Proton Calorimetry/Experimental Runs/2018/Mar22: Difference between revisions

From PBTWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 48: Line 48:
|-
|-


| 1 || f/5 || 1s || Low || colspan="3" | N/A || Background || Background run with ND filter
|| <div class="image150px" style="text-align: center;">
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/MedAustron_Mar2018/Nikon/DSC_0001.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/MedAustron_Mar2018/Nikon/DSC_0001.jpg]
</div>
|| <div class="image150px" style="text-align: center;">
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/MedAustron_Mar2018/EPIX/Run0001.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/MedAustron_Mar2018/EPIX/Run0001.jpg]
</div>
|-
| 2 || f/5 || 1s || Low || 100% || 1.2 x 1e10/5s || 80.5 || Bragg peak || ND filter in - not enough light passes through filter. Remove! DSLR slightly saturated. 10 spills used for this run.
|| <div class="image150px" style="text-align: center;">
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/MedAustron_Mar2018/Nikon/DSC_0002.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/MedAustron_Mar2018/Nikon/DSC_0002.jpg]
</div>
|| <div class="image150px" style="text-align: center;">
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/MedAustron_Mar2018/EPIX/Run0002.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/MedAustron_Mar2018/EPIX/Run0002.jpg]
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/MedAustron_Mar2018/EPIX/Run0002.gif Animation]
</div>
|-





Revision as of 01:57, 21 September 2018

Beam tests of the UCL plastic scintillator range calorimeter.

Night of Thursday 22th to Friday 23rd of March 2018

Pre-data Taking Information and Notes

  • Data was acquired with the DSLR camera and the ISDI CMOS sensor.
  • The average background value per pixel for the sensor is ~ 2500. The sensor pixel dimensions are 1030 x 1536. The sensor frame rate was always set to 25 frames per second.
  • NO optical grease between scintillator and sensor.
  • Beam is extracted over 5 seconds (one spill). The full capacity (100%) of the synchrotron ring is 1.2x1e10 protons.
  • Scintillator stack arrangement: 15 x 2 mm, 15 x 2.6 mm and 1 x 3 mm sheets used: In ascending order from the back of the scintillator to the front/beam. Except for sheet number 21 (2mm) which is placed at the end (away end from the beam) of the stack because of a crack in the sheet. Sheet numbers from front to end (in beam direction): 31,25,24,23,20,16,12,11,9,8,6,5,4,3,2,1,30,29,28,27,26,22,19,18,17,15,14,13,10,7,21
  • Folder for DSLR images: digiCamControl/20180322_medaustron
  • Note: When powering on PC, sensor has to be connected, but unpowered!
  • 16383 = saturation pixel number (16 bit x 1 tap)
  • Sensor images are always saved as a sequence of 21 Tiff images
  • Folder for sensor images: XCAP/data/20180322_medaustron

IMG_3067.jpg

IMG_3057.jpg

Data Taking

DSLR Sensor Beam Info Thumbnails
Run number Aperture Exposure (s) Full Well Mode Intensity (%) Number of particles/s Beam Energy (MeV) Test type Comments DSLR Sensor
1 f/5 1s Low N/A Background Background run with ND filter

DSC_0001.jpg

Run0001.jpg

2 f/5 1s Low 100% 1.2 x 1e10/5s 80.5 Bragg peak ND filter in - not enough light passes through filter. Remove! DSLR slightly saturated. 10 spills used for this run.

DSC_0002.jpg



End of range telescope tests.