Proton Calorimetry/Experimental Runs/2024/UCLHQuARC: Difference between revisions

From PBTWiki
Jump to navigation Jump to search
Line 3: Line 3:
*Stack B:  
*Stack B:  
*DAQ using ** FPGA and Raspberry Pi (running web GUI).
*DAQ using ** FPGA and Raspberry Pi (running web GUI).
==Experiment Notes ==
Stack A + NUC PC.
Beam current nominally 1nA.
Range in eye tissue of 60MeV: 29.3mm
Degrader WET given in eye tissue water equivalent thickness.
Multiply degrader WET by 1.05 to get actual WET.
For modulator measurements, '''Range''' is 29.3mm minus degrader.
'''Modulation''' is eye tissue WET difference between '''Range''' and minimum range.
{| class="wikitable"
!Run number
!Beam Energy (MeV)
!Dose Rate (MU/min)
!Estimated Range in Eye (mm)
!Spot size (mm, FWHM)
!Degrader Eye WET (mm)
!DDC232 FSR (pC)
!DDC232 Integration Time (us)
!Measurements
!Comments
!Photodiode QB Fit
!Replay Fit
|-
|0  || Background || - || - || N/A ||  - || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|1  || 60 || 20 || 29.3 || 34 || 0 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|2  || 60 || 20 || 29.3 || 34 || 0 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|3  || 60 || 20 || 29.3 || 34 || 0 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|4  || 60 || 20 || 28.3 || 34 || 1 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|5  || 60 || 20 || 27.3 || 34 || 2 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|6  || 60 || 20 || 26.3 || 34 || 3 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|7  || 60 || 20 || 25.3 || 34 || 4 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|8  || 60 || 20 || 24.3 || 34 || 5 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|9  || 60 || 20 || 23.3 || 34 || 6 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|10  || 60 || 20 || 22.3 || 34 || 7 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|11  || 60 || 20 || 21.3 || 34 || 8 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|12  || 60 || 20 || 20.3 || 34 || 9 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|13  || 60 || 20 || 19.3 || 34 || 10 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|14  || 60 || 20 || 17.3 || 34 || 12 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|15  || 60 || 20 || 15.3 || 34 || 14 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|16  || 60 || 20 || 13.3 || 34 || 16 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|17  || 60 || 20 || 11.3 || 34 || 18 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|18  || 60 || 20 || 9.3 || 34 || 20 || 12.5 || 170 (constant) || 30,000 ||  || N/A ||
|-
|19  || 60 || 20 || 29.3 || 34 || 0 || 12.5 || 170 (constant) || 100,000 ||  || N/A ||
|-
|20  || 60 || 20 || 28.8 || 34 || 0.6 || 12.5 || 170 (constant) || 100,000 || Modulator reference: 292/92. Range (mm eye tissue): 28.8; Modulation (mm eye tissue): 14.1  || N/A ||
|-
|21  || 60 || 20 || 27.3 || 34 || 2 || 12.5 || 170 (constant) || 100,000 || Modulator reference: 132/19. Range (mm eye tissue): 27.3; Modulation (mm eye tissue): 27.3 || N/A ||
|-
|22  || 60 || 20 || 29.3 || 34 || 0 || 12.5 || 170 (constant) || 100,000 || Modulator reference: 769/02. Range (mm eye tissue): 29.3; Modulation (mm eye tissue): 29.3 || N/A ||
|-
|23  || 60 || 20 || 23.9 || 34 || 5.4 || 12.5 || 170 (constant) || 100,000 || Modulator reference: 172/91. Range (mm eye tissue): 23.9; Modulation (mm eye tissue): 16.5 || N/A ||
|-
|24  || 60 || 20 || 5.6 || 34 || 23.8 || 12.5 || 170 (constant) || 100,000 || Modulator reference: 770/02. Range (mm eye tissue): 5.6; Modulation (mm eye tissue): 5.6 || N/A ||
|-
|25  || 60 || 50 || 29.3 || 34 || 0 || 50 || 170 (constant) || 30,000 ||  || N/A ||
|-
|}

Revision as of 15:31, 10 April 2024

Detector Equipment

  • Stack A:
  • Stack B:
  • DAQ using ** FPGA and Raspberry Pi (running web GUI).


Experiment Notes

Stack A + NUC PC.

Beam current nominally 1nA.

Range in eye tissue of 60MeV: 29.3mm

Degrader WET given in eye tissue water equivalent thickness. Multiply degrader WET by 1.05 to get actual WET.

For modulator measurements, Range is 29.3mm minus degrader. Modulation is eye tissue WET difference between Range and minimum range.

Run number Beam Energy (MeV) Dose Rate (MU/min) Estimated Range in Eye (mm) Spot size (mm, FWHM) Degrader Eye WET (mm) DDC232 FSR (pC) DDC232 Integration Time (us) Measurements Comments Photodiode QB Fit Replay Fit
0 Background - - N/A - 12.5 170 (constant) 30,000 N/A
1 60 20 29.3 34 0 12.5 170 (constant) 30,000 N/A
2 60 20 29.3 34 0 12.5 170 (constant) 30,000 N/A
3 60 20 29.3 34 0 12.5 170 (constant) 30,000 N/A
4 60 20 28.3 34 1 12.5 170 (constant) 30,000 N/A
5 60 20 27.3 34 2 12.5 170 (constant) 30,000 N/A
6 60 20 26.3 34 3 12.5 170 (constant) 30,000 N/A
7 60 20 25.3 34 4 12.5 170 (constant) 30,000 N/A
8 60 20 24.3 34 5 12.5 170 (constant) 30,000 N/A
9 60 20 23.3 34 6 12.5 170 (constant) 30,000 N/A
10 60 20 22.3 34 7 12.5 170 (constant) 30,000 N/A
11 60 20 21.3 34 8 12.5 170 (constant) 30,000 N/A
12 60 20 20.3 34 9 12.5 170 (constant) 30,000 N/A
13 60 20 19.3 34 10 12.5 170 (constant) 30,000 N/A
14 60 20 17.3 34 12 12.5 170 (constant) 30,000 N/A
15 60 20 15.3 34 14 12.5 170 (constant) 30,000 N/A
16 60 20 13.3 34 16 12.5 170 (constant) 30,000 N/A
17 60 20 11.3 34 18 12.5 170 (constant) 30,000 N/A
18 60 20 9.3 34 20 12.5 170 (constant) 30,000 N/A
19 60 20 29.3 34 0 12.5 170 (constant) 100,000 N/A
20 60 20 28.8 34 0.6 12.5 170 (constant) 100,000 Modulator reference: 292/92. Range (mm eye tissue): 28.8; Modulation (mm eye tissue): 14.1 N/A
21 60 20 27.3 34 2 12.5 170 (constant) 100,000 Modulator reference: 132/19. Range (mm eye tissue): 27.3; Modulation (mm eye tissue): 27.3 N/A
22 60 20 29.3 34 0 12.5 170 (constant) 100,000 Modulator reference: 769/02. Range (mm eye tissue): 29.3; Modulation (mm eye tissue): 29.3 N/A
23 60 20 23.9 34 5.4 12.5 170 (constant) 100,000 Modulator reference: 172/91. Range (mm eye tissue): 23.9; Modulation (mm eye tissue): 16.5 N/A
24 60 20 5.6 34 23.8 12.5 170 (constant) 100,000 Modulator reference: 770/02. Range (mm eye tissue): 5.6; Modulation (mm eye tissue): 5.6 N/A
25 60 50 29.3 34 0 50 170 (constant) 30,000 N/A