2025/08/05: Difference between revisions
Jump to navigation
Jump to search
(Created page with "== Minutes for UCL Proton Beam Therapy Group Meetings, 29th July == === Present === '''Joe Bateman''', '''Sonia Escribano''', '''Simon Jolly'''") |
|||
Line 3: | Line 3: | ||
=== Present === | === Present === | ||
'''Joe Bateman''', '''Sonia Escribano''', '''Simon Jolly''' | '''Joe Bateman''', '''Sonia Escribano''', '''Febian''', '''Simon Jolly''' | ||
=== [[ELogs/SimonJolly|Simon Jolly]] === | |||
* Ordered two starter kits for kria | |||
* Potentially do FLASH calibration of module 2 in addition to conv calibration of all 3 at Trento | |||
=== [[ELogs/SoniaEscribano|Sonia Escribano]] === | |||
* Tested revF boards - 4 of the 8 working boards - fully working (other 4 full of soldering residue so can’t connect PDs) | |||
* 4 working boards now assembled within 3-module QuARC and working | |||
* Sorted cables for Trento beam test and spares | |||
* Will need to do shoot throughs of first two modules since have new boards - no changes to module 3 but will have another shoot through | |||
* One zmod board not working | |||
* Two of the fpga’s only work with 5V | |||
* For beam shift QuARC measurements do difference over sum for position measurements | |||
* Run 66 overwritten with background data |
Revision as of 09:03, 6 October 2025
Minutes for UCL Proton Beam Therapy Group Meetings, 29th July
Present
Joe Bateman, Sonia Escribano, Febian, Simon Jolly
Simon Jolly
- Ordered two starter kits for kria
- Potentially do FLASH calibration of module 2 in addition to conv calibration of all 3 at Trento
Sonia Escribano
- Tested revF boards - 4 of the 8 working boards - fully working (other 4 full of soldering residue so can’t connect PDs)
- 4 working boards now assembled within 3-module QuARC and working
- Sorted cables for Trento beam test and spares
- Will need to do shoot throughs of first two modules since have new boards - no changes to module 3 but will have another shoot through
- One zmod board not working
- Two of the fpga’s only work with 5V
- For beam shift QuARC measurements do difference over sum for position measurements
- Run 66 overwritten with background data