A reading list for QA devices

Articles on Detectors

  • "Dosimetry for ion beam therapy" (2010)

This paper looks at current methods and standards for PBT dosimetry. It also gives advantages and disadvantages of each method. This paper was particularly useful for looking at film dosimetry and ionization chambers. https://iopscience.iop.org/article/10.1088/0031-9155/55/21/R01/meta

  • "A critical review of the practices of proton daily quality assurance programs" (2021)

https://dx.doi.org/10.21037/TRO-21-11

Commercially Available Devices

IBA Dosimetry

  • Lynx/Sphinx
    • Usage: Daily QA, Monthly QA, Annual QA, Commissioning
    • For performing spot size and position measurements and numerous energies.
    • 0.4 mm thick and 300 x 300 mm^2 Ga- based plastic scintillator screen, with mirror and CCD camera.
    • Technical specs:
      • Pixel size: 4.65 x 4.65 um²
      • Effective spacial resolution: 0.5 mm
      • Video camera: 1024 x 1024, 12-bit resolution
      • Sampling time: 7 images / s
      • Dynamic range can be modified through CCD camera gain settings and lens aperture.
      • Total dimensions 360 x 370 x 600 mm
    • Sphinx phantom consists of wedges for range consistency checks for 4 energies (no direct range measurement) and absolute dose measurements with PPC05 Plane Parallel Chamber.
    • Used with myQA machine software: myQA Machines Software




  • myQA Phoenix
    • Usage: Monthly QA, Annual QA, Commissioning
    • High-resolution amorphous silicon digital detector array.
    • Capable of single shot and movie mode measurements.
    • Technical specs:
      • 2048 × 2048 (4,194,304) pixels.
      • 0.2mm resolution.
      • 1 frame/sec with 65536 counts/frame sampling frequency.


  • myQA MatriXX AiR
    • Usage: PSQA
    • Wireless ionisation chamber array used for planar verification of of patient treatment plans.
    • WiFi and battery operation mode available.
    • Technical specs:
      • 1521 air-vented ICs
      • 25 × 25 cm² area
      • 6.5 mm center-to-center resolution
      • Minimum sampling time of 20 ms
      • Detector dimensions: 57.6x32x4cm
      • 7.3kg weight (inc battery)
    • Reads out 2D dose distribution at different depths and gantry angles using miniPhantom R & Gantry Sensor+.
    • Compatible with protons and carbon ions.
    • conformalFLASH ready
    • Compatible with DynamicArc (proton arc therapy) using gantry angle sensor and rotating holder.




  • Giraffe
    • Usage: Monthly QA, Annual QA, Commissioning
  • Zebra
    • Usage: Monthly QA, Annual QA, Commissioning
  • StingRay
    • Usage: Monthly QA, Annual QA, Commissioning

PTW

Sun Nuclear

DET.TEC.TOR

Logos Systems

MedScint