Proton Calorimetry/Future Work

From PBTWiki
Jump to navigation Jump to search

Personnel

  • UCL: Simon Jolly (SJ), Joseph Bateman (JB), Febian (F), Matt Warren (MW), Harry Barnett (HB).
  • Bari (Ba): Raffaella Radogna (RR).
  • Oxford University (Ox): Pete Hastings (PH), Mark Jones (MJ).
  • Imperial College (IC): Andy Rose (AR), Duncan Parker (DP), Munir Saleh (MS).
  • Heidelberg (He): Blake Leverington (BL), Julian Horn (JH).

Design Goals

  • Real-time measurements of PBT beam range with sub-millimetre precision using plastic scintillator range telescope.
  • Real-time measurement of PBT beam spot size/position with sub-millimetre precision using scintillating fibre arrays.
  • Coverage of “clinically relevant” transverse scanning area and range:
    • 230 MeV/330 mm range.
    • 5 cm × 5 cm scanning area.
  • Integration into single portable detector enclosure with single power and network inputs and nozzle mount.
  • Full system control and display with web-based GUI from on-board DAQ.

Current Status (October 2025)

  • QuARC tested in Trento:
    • Calibration stable.
    • Range measurements repeatable.
    • Full GUI control.
  • Fibre array prototype tested in Trento:
    • Analogue Hamamatsu photodiode arrays and driver circuits.
    • NI-driven DAQ.
    • Fibre arrays assembled in Italy with 500 micron scintillating fibres.
    • Repeatable, low noise measurements of beam position and profile at multiple positions and sizes.

QuARC Development

  • Reduce scintillator sheet thickness/photodiode spacing to 2.5 mm.
  • Improve manufacturing process and surface finish.
  • Move from 2 to 4-sided photodiode readout.
  • Replace individual photodiodes with 16-element arrays.
  • Daisy-chain electronics around 4 sides of single module.
  • Increase scintillator area to 150 × 150 mm.

Scintillator Development

  • Existing scintillator sheets need machining on Datron to improve quality and reduce thickness to 2.5 mm: HB.
  • Superior quality scintillator blanks needed from Nuvia: HB and SJ to liaise with Hana Buresova.
  • Move to larger area scintillator sheets once 100 mm sheet production refined.

Scintillator Modules

  • Redesign scintillator stack holders for 4-sided readout: HB.
  • Individual modules must be self-contained and pre-cabled on 3 sides with only top-mounted USB-C I/O connector needing to be connected when assembling.

Photodiodes and Electronics

  • Front-end DDC boards need to shrink to accommodate Hamamatsu S12362 16-element arrays with 2.5 mm pitch: PH.
  • Replace vertical USB-C connectors with horizontal and move to upstream end to match new 4-sided setup: PH to liaise with HB.

DAQ and GUI

  • Replace Raspberry Pi/USB104 with QuADProBe Kria (see below): F.
  • Live GUI frequently unresponsive, needs resolving: JB.
  • Replace PapaParse with direct JSON input: JB.
  • Ensure post processing and replay fully functional to review previous beam tests: JB.

Mechanical Design

QuADProBe Development

  • 250 micron square scintillating fibre arrays: RR.
  • Hamamatsu digital S17285 photodiode array readout.
  • Kria-based DAQ

Calibration and data

Is it possible to calibrate the detector without a proton beam?

  • Maybe an e- beam beta source?
  • Modify the geant4 simulation to verify what E will we need so there is no quenching?
  • Can we obtain average results to mitigate position sensitivity?

GUI

  • Implement any options needed that have user input
  • Replay and post-processing options
  • Version with full live detector control ➾ reset and relaunch options included
  • Check current varian control system: what does if offer and how do they control the detector? ➾ Check with Clatterbridge, Allison and A. Mazal

Mechanical design

  • For 245 MeV, is it doable to do 1 module of 32 PD with the 2.8mm thick and then 3 modules of 32 PD with scintillators 3mm thick?
  • How can we design the modules so they can attach and detach easily? ➾ We need a robust design but not rigid
  • Power distribution?
  • Beam windows? Light tight?

Future beam tests

  • Beam test where we check the shape and position of the BP (scanning field)
  • Real treatment plans
  • Testing with real gantry mount