Proton Calorimetry/Future Work
Jump to navigation
Jump to search
Design Goals
- Real-time measurements of PBT beam range with sub-millimetre precision using plastic scintillator range telescope.
- Real-time measurement of PBT beam spot size/position with sub-millimetre precision using scintillating fibre arrays.
- Coverage of “clinically relevant” transverse scanning area and range:
- 230 MeV/330 mm range.
- 5 cm× 5 cm scanning area.
- Integration into single portable detector enclosure with single power and network inputs and nozzle mount.
- Full system control and display with web-based GUI from on-board DAQ.
Current Status (October 2025)
Scintillators and PD
- Scintillators uniform in thickness (3mm) and flatness.
- Machine block clear polished sheets.
- For 245 MeV, is it doable to do 1 module of 32 PD with the 2.8mm thick and then 3 modules of 32 PD with scintillators 3mm thick?
- For low energy just 1 module of 2.8mm
- Photodiodes are OK, there are no issues in the coupling.
DAQ & Electronics
Redesign of the analog circuit. The boards need a revision.
- We want everything to connect via usb-C (currently there is only one usb-C cable that does not work properly and it is a very long one). Did Saad discuss this with Marco? What was the outcome of the lab tests?
- We need all the power to come just from FPGA (5V, ~2A) ➾ the voltage regulator chips need to go.
- Long usb-C between the daugther board and the DDC will be removed, and the two can connect directly [daughter board - bridge usb C board - (usb C cable) - bridge board - DDC board]
- FPGA from Saad had some glitches that need to be resolved. Do we see those with Nexus as well as the usb104? ➾ On board fitting needs to be implemented
Calibration and data
Is it possible to calibrate the detector without a proton beam?
- Maybe an e- beam beta source?
- Modify the geant4 simulation to verify what E will we need so there is no quenching?
- Can we obtain average results to mitigate position sensitivity?
GUI
- Implement any options needed that have user input
- Replay and post-processing options
- Version with full live detector control ➾ reset and relaunch options included
- Check current varian control system: what does if offer and how do they control the detector? ➾ Check with Clatterbridge, Allison and A. Mazal
Mechanical design
- For 245 MeV, is it doable to do 1 module of 32 PD with the 2.8mm thick and then 3 modules of 32 PD with scintillators 3mm thick?
- How can we design the modules so they can attach and detach easily? ➾ We need a robust design but not rigid
- Power distribution?
- Beam windows? Light tight?
Future beam tests
- Beam test where we check the shape and position of the BP (scanning field)
- Real treatment plans
- Testing with real gantry mount