Proton Calorimetry/Experimental Runs/2018/Aug16

From PBTWiki
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

Beam tests of the UCL plastic scintillator range calorimeter.

Night of 15th to 16th of August 2018

Pre-data Taking Information and Notes

Sheet ordering from front to end (in beam direction): 31,25,24,23,20,16,12,11,9,8,6,5,4,3,2,1,30,29,28,27,26,22,19,18,17,15,14,13,10,7,21

100% beam intensity is about 10e6 particles per second

The MedAustron Ripple filter used in this experiment has a WET of 3mm.

The sensor pixel dimensions are 1030 x 1536 which has to be set manually


Serial Terminal instructions for ISDI sensor:

In Controls/Options: Make sure "Send string with CR" is selected. Should be default.

In Controls/Setup: Select Serial Port: Enabled. Baud rate has to be 115200

Set low full well mode using W000300000

Set high full well mode using W000300004

Read full well mode using R0003



Data Taking

1 hour of beam time: 30min setup, 30 min measurement time

Mistake during data taking: Full well mode should have been "high" in order to have a higher saturation threshold!

Run number Full Well Mode Beam Energy (MeV/u) Intensity Degrader Test type Comments Thumbnail
0 low 0 10% background two ripple filter in! have a WET of 3.0mm until stated otherwise

Run000.jpg

1 low 184.5 10% energy scan

Run001.jpg

2 low 184.5 10% energy scan repeat

Run002.jpg

3 low 184.5 10% energy scan repeat

Run003.jpg

4 low 400 10% calibration RF calibration shootthrough with ripple filters!

Run004.jpg

5 low 400 10% calibration RF repeat

Run005.jpg

6 low 154.5 10% energy scan

Run006.jpg

7 low 146.5 10% energy scan

Run007.jpg

8 low 138.3 10% energy scan

Run008.jpg

9 low 121.1 10% energy scan

Run009.jpg

10 low 120 10% energy scan

Run010.jpg

11 low 0 10% background ripple filter out! Background

Run011.jpg

12 low 400 10% calibration calibration shootthrough

Run012.jpg

13 low 184.5 10% energy scan

Run013.jpg

14 low 184.5 10% energy scan repeat

Run014.jpg

15 low 154.5 10% energy scan

Run015.jpg

16 low 146.5 10% energy scan

Run016.jpg

17 low 138.3 10% energy scan

Run017.jpg

18 low 121.1 10% energy scan

Run018.jpg

19 low 120 10% energy scan

Run019.jpg

20 low 184.5 10% radiation damage check for radiation damage

Run020.jpg

21 low 184.5 10% radiation damage repeat

Run021.jpg