Proton Calorimetry/Experimental Runs/2018/Mar8-9: Difference between revisions

From PBTWiki
Jump to navigation Jump to search
Line 38: Line 38:
!Sensor Thumbnail
!Sensor Thumbnail
|-
|-
| 1 || f/10 || 2 s || N/A || N/A || N/A || N/A || N/A || Beam off. Focus check for DSLR. Lid open. Low light. ||  
| 1 || f/10 || 2 s || N/A || N/A || N/A || N/A || N/A || Beam off, lid open. Focus check for DSLR. Low light. ||  
<div class="image150px" style="text-align: center;">
<div class="image150px" style="text-align: center;">
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0001.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0001.jpg]
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0001.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0001.jpg]
Line 45: Line 45:
|| N/A
|| N/A
|-
|-
| 2 || f/10 || 5 s || N/A || N/A || N/A || N/A || N/A || Beam off. Focus check for DSLR. Lid open. Low light. ||  
| 2 || f/10 || 5 s || N/A || N/A || N/A || N/A || N/A || Beam off, lid open. Focus check for DSLR. Low light. ||  
<div class="image150px" style="text-align: center;">
<div class="image150px" style="text-align: center;">
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0002.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0002.jpg]
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0002.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0002.jpg]
Line 51: Line 51:
</div>  
</div>  
|| N/A
|| N/A
|-
| 3 || f/10 || 20 s || N/A || N/A || N/A || N/A || N/A || Beam off, lid open. Focus check for DSLR. Low light. ||
<div class="image150px" style="text-align: center;">
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0003.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0003.jpg]
<br />
</div>
|| N/A
|-
| 4 || f/10 || 20 s || N/A || N/A || N/A || N/A || N/A || Beam off, lid open. Focus check for DSLR. Focus - okay. ||
<div class="image150px" style="text-align: center;">
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0004.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0004.jpg]
<br />
</div>
|| N/A
|-
| 5 || f/5 || 5 s || N/A || N/A || N/A || N/A || N/A || Beam off, lid open. Focus check and background picture with white paper for DSLR. ||
<div class="image150px" style="text-align: center;">
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0005.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0005.jpg]
<br />
</div>
|| N/A
|-
| 6 || f/10 || 5 s || Low || 42 frames || N/A || N/A || N/A || Beam off, lid closed, lights off. Background run. ||
<div class="image150px" style="text-align: center;">
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0006.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/images/Birmingham_Mar2018/Nikon/20180308_bham/DSC_0006.jpg]
<br />
</div>
|| N/A
|-

Revision as of 12:39, 12 March 2018

Beam tests of the UCL plastic scintillator range calorimeter.

Thursday 8th of March 2018

Pre-data Taking Information and Notes

  • The tests were carried out with an energy of 36 MeV. A Bragg Peak measurement was carried out to confirm the exact beam energy.
  • Collimators used: 2 mm and 20 mm diameter.
  • PMMA sheets used: 3 mm.
  • The average background value per pixel for the sensor is ~ 2500. The sensor pixel dimensions are 1030 x 1536. The sensor frame rate was always set to 25 frames per second.
  • The DSLR ISO was always set to 200.
  • 6 x 3 mm scintillator sheets were used. The scintillator stack was set up in the following order, starting with the first sheet facing the beam:
    • 31, 32, 33, 34, 35, 48
  • The scintillator stack was coupled to the sensor using optical gel.
  • The dose received (as shown on the monitor in the control room) during the full day of measurements was:
    • With a 2 mm collimator: 746.9 uC
    • With a 20 mm collimator: 417.7 uC
    • With a 2 mm collimator + 3 mm of PMMA: 251.5 uC
  • To calculate the actual dose delivered to the scintillator:
    • / 160 -> to convert to proton current
    • / (50/ collimator diameter)^2 -> to account for reduction in area due to the collimator
  • The directory containing the sensor data is:

/unix/pbt/data/birmingham/mar_2018/EPIX/XCAP/data/20180308_bham

  • The directory containing the DSLR data is:

/unix/pbt/data/birmingham/mar_2018/Nikon/20180308_bham

Data Taking

Run number DSLR Aperture DSLR Exposure Sensor Full Well Mode Sensor Sequence Beam Energy Current on Monitor Collimator Diameter Comments DSLR Thumbnail Sensor Thumbnail
1 f/10 2 s N/A N/A N/A N/A N/A Beam off, lid open. Focus check for DSLR. Low light.

DSC_0001.jpg

N/A
2 f/10 5 s N/A N/A N/A N/A N/A Beam off, lid open. Focus check for DSLR. Low light.

DSC_0002.jpg

N/A
3 f/10 20 s N/A N/A N/A N/A N/A Beam off, lid open. Focus check for DSLR. Low light.

DSC_0003.jpg

N/A
4 f/10 20 s N/A N/A N/A N/A N/A Beam off, lid open. Focus check for DSLR. Focus - okay.

DSC_0004.jpg

N/A
5 f/5 5 s N/A N/A N/A N/A N/A Beam off, lid open. Focus check and background picture with white paper for DSLR.

DSC_0005.jpg

N/A
6 f/10 5 s Low 42 frames N/A N/A N/A Beam off, lid closed, lights off. Background run.

DSC_0006.jpg

N/A