Proton Calorimetry/Experimental Runs/2019/Mar26: Difference between revisions
SaadShaikh (talk | contribs)  (Created page with "Pulse Generator Settings * Frequency: 10 MHz * Delay: 0 ps * Width: 10 ns * LeadE: 5 ns * TrailE = LeadE * High = 60 mV * Low = 0 mV  Measured the number of triggers for 100 seco...")  | 
				SaadShaikh (talk | contribs)  No edit summary  | 
				||
| Line 1: | Line 1: | ||
Pulse Generator Settings  | '''Pulse Generator Settings'''  | ||
* Frequency: 10 MHz  | * Frequency: 10 MHz  | ||
* Delay: 0 ps  | * Delay: 0 ps  | ||
| Line 11: | Line 11: | ||
50 ns/div is the default timebase.    | 50 ns/div is the default timebase.    | ||
Real-time Mode, 10MHz, Amplitude Measurement  | '''Real-time Mode, 10MHz, Amplitude Measurement'''  | ||
* 100 ns/div = 42956 -> 2.33 ms  | * 100 ns/div = 42956 -> 2.33 ms  | ||
* 50 ns/div =  43233 -> 2.31 ms  | * 50 ns/div =  43233 -> 2.31 ms  | ||
| Line 19: | Line 19: | ||
Value in brackets for DTrig runs is the on-screen displayed average trigger interval.  | Value in brackets for DTrig runs is the on-screen displayed average trigger interval.  | ||
Real-time Mode, 10MHz, DTrig Measurement    | '''Real-time Mode, 10MHz, DTrig Measurement'''  | ||
* 100 ns/div = 45061 -> 2.22 ms (2.2235 ms)  | * 100 ns/div = 45061 -> 2.22 ms (2.2235 ms)  | ||
* 50 ns/div = 45511 -> 2.20 ms (2.2011 ms)  | * 50 ns/div = 45511 -> 2.20 ms (2.2011 ms)  | ||
| Line 27: | Line 27: | ||
Sequence mode used 50000 triggers per file: don’t measure number of triggers here, just the DTrig time.  | Sequence mode used 50000 triggers per file: don’t measure number of triggers here, just the DTrig time.  | ||
Sequence Mode, 10MHz, DTrig Measurement  | '''Sequence Mode, 10MHz, DTrig Measurement'''  | ||
* 100 ns/div = 2.58057 microseconds  | * 100 ns/div = 2.58057 microseconds  | ||
* 50 ns/div = 1.61644 microseconds  | * 50 ns/div = 1.61644 microseconds  | ||
Latest revision as of 11:22, 16 November 2020
Pulse Generator Settings
- Frequency: 10 MHz
 - Delay: 0 ps
 - Width: 10 ns
 - LeadE: 5 ns
 - TrailE = LeadE
 - High = 60 mV
 - Low = 0 mV
 
Measured the number of triggers for 100 seconds. Divide by 100 and then inverse to find trigger time interval. 50 ns/div is the default timebase.
Real-time Mode, 10MHz, Amplitude Measurement
- 100 ns/div = 42956 -> 2.33 ms
 - 50 ns/div = 43233 -> 2.31 ms
 - 20 ns/div = 44276 -> 2.26 ms
 - 10 ns/div = 44120 -> 2.27 ms
 
Value in brackets for DTrig runs is the on-screen displayed average trigger interval.
Real-time Mode, 10MHz, DTrig Measurement
- 100 ns/div = 45061 -> 2.22 ms (2.2235 ms)
 - 50 ns/div = 45511 -> 2.20 ms (2.2011 ms)
 - 20 ns/div = 46260 -> 2.16 ms (2.1669 ms)
 - 10 ns/div = 46249 -> 2.16 ms (2.1705 ms)
 
Sequence mode used 50000 triggers per file: don’t measure number of triggers here, just the DTrig time.
Sequence Mode, 10MHz, DTrig Measurement
- 100 ns/div = 2.58057 microseconds
 - 50 ns/div = 1.61644 microseconds
 - 20 ns/div = 1.0426 microseconds
 - 10 ns/div = 949 ns
 - 5 ns/div = 906.9 ns
 - 2 ns/div = 914.63 ns