3 night shifts with range calorimeter + SDLR
Afternoon test at UCLH with photodiodes and CMOS sesnor.


== Experiment Equipment ==
== Experiment Equipment ==
Line 10: Line 10:
|-
|-
| style="text-align: center;" | Control Laptop x3  
| style="text-align: center;" | Control Laptop x3  
| style="text-align: center;" | 2x remote control for sensor and DSLR, 1 for notes
| style="text-align: center;" | 2x remote control for sensor and FPGA, 1 for notes/web GUI.
|-
|-
| style="text-align: center;" | Ethernet Cable x 2
| style="text-align: center;" | Ethernet Cable x 3
| style="text-align: center;" | To connect DAQ laptop to network in the experimental room, network hub to network in control room.
| style="text-align: center;" | To connect DAQ laptops to network in the experimental room, network hub to network in control room.
|-
|-
| style="text-align: center;" | Portable Enclosure
| style="text-align: center;" | Portable Enclosure
| style="text-align: center;" | Modified Big Zarges Waterproof Wheeled Equipment Case.  
| style="text-align: center;" | Modified Big Zarges Waterproof Wheeled Equipment Case.  
Features mount for scintillator and DSLR, front and back openings for beam, patch panel with ports for SHV, BNC, SMA, USB, Camera Link cables. <br/>  
Features mount for scintillator stack, front and back openings for beam, patch panel with ports for SHV, BNC, SMA, USB, Camera Link cables. <br/>  
Hole for sensor powering. <br/>
Hole for sensor powering. <br/>
Mylar windows with Al support mounted to internal face: approximately light-tight.
Mylar windows with Al support mounted to internal face: approximately light-tight.
|-
|-
| style="text-align: center;" | Scintillator stack
| style="text-align: center;" | Scintillator stack
| style="text-align: center;" | 14 x 2 mm, 15 x 2.6 mm and 20 x 3 mm sheets in ascending order from the front/beam of the end of the scintillator. Sheet 21 not used. Sheet numbers from front to end (in beam direction): 30,29,28,27,26,22,19,18,17,15,14,13,10,7,25,24,23,20,16,12,11,9,8,6,5,4,3,2,1,31,32,33,34,35,36,37,38,39,40,41,42,43,44,45,46,47,48,49,50.
| style="text-align: center;" | X x 10 mm, Y x 10 mm and Z x (approx) 2.8 mm sheets in ascending order from the front/beam of the end of the scintillator. Sheet numbers from front to end (in beam direction gantry angle 90°): 25, 31, 50, 52, 34, 44, 53, 47, 33, 54, 46, 42, 55, 45, 38, 36, 43, 32, 48, 41, 8, 39, 40, 37, 35, 49, 9, 2, 1, 51, 11, 4, 3, 23, 6. Full stack thickness measurements (in mm): 98.10, 98.08, 97.93, 98.49.
|-
|-
| style="text-align: center;" | ISDI CMOS sensor
| style="text-align: center;" | ISDI CMOS sensor
Line 30: Line 30:
| style="text-align: center;" | Connected to DAQ Laptop via USB cable.  
| style="text-align: center;" | Connected to DAQ Laptop via USB cable.  
|-
|-
| style="text-align: center;" | DAQ laptop
| style="text-align: center;" | DAQ laptop x2
| style="text-align: center;" | Controls sensor aquisition and DSLR aquisition.
| style="text-align: center;" | Control sensor and photodiode acquisition.
|-
| style="text-align: center;" | Nexys Video FPGA development board.
| style="text-align: center;" | For interfacing between DDC232 and PC.
|-
| style="text-align: center;" | Rev. 2 Texas Instruments DDC232 custom circuit board (x2)
| style="text-align: center;" | Each housing 16x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets. No optical gel and aligned by eye to scintillator sheets. Photodiodes do not cover sheets 25, 23 and 6.
|-
|-
| style="text-align: center;" | Gloves
| style="text-align: center;" | Gloves
| style="text-align: center;" | For handling scintillator
| style="text-align: center;" | For handling scintillator
|-
| style="text-align: center;" | Digital Caliper
| style="text-align: center;" | For measuring PMMA air gaps
|}
|}
==== Range Detector Experiment ====
==== Schematic ====
<div class="image600px" style="text-align: center;">
<div class="image600px" style="text-align: center;">
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/Heidelberg_Apr2019/HITApr2019_setup.png http://www.hep.ucl.ac.uk/pbt/wikiData/images/Heidelberg_Apr2019/HITApr2019_setup.png]
[http://www.hep.ucl.ac.uk/pbt/wikiData/images/UCLH_Apr2021/UCLH_Setup.png http://www.hep.ucl.ac.uk/pbt/wikiData/images/UCLH_Apr2021/UCLH_Setup.png]
</div>
</div>


Line 66: Line 69:
*This setting makes sure that the buffer is filled with a new frame as they come in. If this option is not selected the sensor and software will not be synchronised and we *will record a lot of pitch black images.
*This setting makes sure that the buffer is filled with a new frame as they come in. If this option is not selected the sensor and software will not be synchronised and we *will record a lot of pitch black images.
*Select Capture/Live Options//Live Mode/Live Video: Continuous All buffer capture
*Select Capture/Live Options//Live Mode/Live Video: Continuous All buffer capture
=== DDC232 Instructions ===
*Bash script <code>prog.sh</code> executes <code>FTDI.cpp</code> and <code>livePlot.cpp</code>. Execute with <code>. prog.sh 2 350 170 0 50 true Data/capture.txt</code> in the directory <code>Photodiodes</code>.
**First argument is the number of DDC232s.
**Second argument is the full-scale range (FSR) in pC. Choices are: 12.5, 50, 100, 150, 200, 250, 300, 350.
**Third argument is the integration time in us. Range is 170-1455. Set in multiples of 5 us.
**Fourth argument is the number of measurements to acquire. Typically acquire 5700 measurements per second. Set to 0 for infinite loop.
**Fifth argument is the rate in Hz to update the file <code>average.csv</code> for ROOT/web GUI plotting.
**Sixth argument is a boolean to choose whether to display the ROOT plot.
**Seventh argument is the filename and destination for the output hexadecimal data.
*After acquisition, close ROOT by choosing File->Quit in the ROOT window and then stopping DAQ (if in infinite loop) with CTRL-C.


== Experiment Plan ==
== Experiment Plan ==


=== 12&ndash;13th April ===
*Shoot-through measurements (back & front) for light calibration and level setting.
# Find convenient beam intensity that does not saturate sensor: lowest clinical beam energy, smallest beam focus
*Range measurements between 70-120 MeV with 1mm steps with CMOS sensor and photodiodes.
# Calibration of detector (highest clinical energy). Shoot from front and back of stack (we have a new entrance window at the back of the stack). Also calibrate at different focus sizes. Also test camera: focus on front of beam stack and centre of beam stack and compare.
*High intensity range measurements between 70-120 MeV with 1mm steps with photodiodes at optimal light level.
# Test focus (beam size) dependence: shoot pencil beam with ~10 cm range at different focus sizes
*Live measurement of 3cm x 3cm x 3cm treatment plan.
# Measure WET of known blocks of plastic to calibrate range telescope using a pencil beam with ~10cm range. Lennart will check if he can find some PMMA slabs whose WET has been accurately measured. Ideally have some slabs between ~1mm and ~5cm WET thickness. We know that there are some PMMA slabs between 1mm and 5cm thickness, measured by Giulia Arico (OMA fellow). These measurements will be used to benchmark Laurent's reconstruction code. Use tightest beam focus.
# Repeat 0.-3. with all available ions (protons, helium, carbon and oxgen)
# Measure SOBP of protons. 10cm deep, 3cm plateau
# Measure high proton energies (up to 250MeV) using PMMA absorber in front of range telescope.
# Measure a couple of oxygen pencil beams (range shift) and SOBP
# Disassemble detector.
 
=== 13&ndash;14th April ===
 
# Shoot-through calibration for Helium and Carbon: 430.1MeV/u Carbon 30cm range, He 220.5MeV/u 30cm range.
## High full-well mode C Intensity: 8*10^7; He Intensity: 8x10^8.
## Low full well mode C Intensity: 3*10^7; He Intensity 3*10^8.
## Rotate detector; repeat.
<!-- -->
# Carbon and Helium SOBP and single energy pencil beam:
## 2 Gy SOBP Carbon, Helium mix keep same intensity here and scale down later in data processing.
## Carbon and Helium Intensities both 8*10^7 (low Helium intensity!).  Both high and low full well modes.  Only measure low full well as sensor starts to come out of saturation.
## Carbon SOBP max range ~9cm (plateau 3cm longitudinal, 3x3cm2 lateral), He SOBP max range ~27 cm.
## Repeat with 220MeV/u pencil beam (C 10cm, He 30 cm).
<!-- -->
# Add PMMA blocks, repeat previous measurement. 24 measurements; 5 changes in room. (5+5+5+2+2cm PMMA blocks, 19cm PMMA in total). Both high and low full well mode.
<!-- -->
# Variable air gap measurements:
## 5cm PMMA, air gap, remaining PMMA absorber; total thickness of 20.8cm. Downstream sheets 5+5+2cm PMMA sheets plus half of unused pairs.
## Vary air gap width: 0mm, 2mm, 5mm, 10mm.
## Vary air gap depth: 1mm, 2mm, 5mm, 10mm.
## Upstream PMMA absorber 5cm; downstream PMMA absorber varies from 12.8cm to 13.7cm.
## 32 measurements, 16 absorber changes.
## For 5mm x 5mm gap, adjust focal spot size.
## Energy 220 MeV/u for both; intensity 8*10^7 for both; use focus setting 3 (8.5mm for Carbon; 8.1mm for Helium).
## Low full well mode for sensor.
<!-- -->
# Half covered 2mm air gap:
## 5cm upstream PMMA, 2mm air gap on left-hand side, 13.6cm downstream PMMA absorber. 18.8cm total PMMA thickness.
## Sheet pulled out, 8mm pulled out from centre, half-centred, 8mm pushed in from centred, all the way in.
## Energy 210 MeV/u for both; intensity 8*10^7 for both; use focus setting 3 (8.5mm for Carbon; 8.1mm for Helium).
## Scan layer with 10x10 spots over 3x3cm area.
## Low full well mode for sensor.
## Take movie with DSLR.
<!-- -->
# Moving platform measurements:
## Assemble PMMA stack on moving platform: upstream PMMA absorber 5cm; 5mm air gap; downstream PMMA absorber 13.3cm.  Total thickness still 18.8cm.
## Gap width 5mm; gap depth 5mm.
## Irradiate with single energy Helium plan: 10x1 spot array (3cm wide) and move platform at the same time. Spot dwell time ~200ms.  Total plan irradiation time: 5s.  Total platform movement: 4cm.
## Take images with sensor: use 100 pixel wide sensor strip and long acquisition.
## Energy 210 MeV/u for both; intensity 8*10^7 for both; use focus setting 3 (8.5mm for Carbon; 8.1mm for Helium).
<!-- -->
# Lung phantom measurements only(!) if enough time. Take lung phantom, place it on height adjustable table, place tumor in beam (180 MeV/u C and He), lift it by 0.5 cm
 
=== 14&ndash;15th April ===
 
# '''Shoot-through calibration for Helium and Carbon: 430.1MeV/u Carbon 30cm range, focus 4.9mm; He 220.5MeV/u 30cm range, focus 3.4mm. 5cm upstream PMMA absorber.'''
## High full-well mode C Intensity: 8*10^7; He Intensity: 8x10^8.
## Low full well mode C Intensity: 3*10^7; He Intensity 3*10^8.
<!-- -->
# '''Lung phantom measurements with tumour.'''
## Align lung phantom with lasers. Lung tumour (about 1cm^3) sits positioned at isocentre.
## Monoenergetic pencil beam: Helium energy 180 MeV/u, Intensity 8*10^7, spot focus 3, Carbon 180MeV/u, intensity 8*10^7, spot focus 3.
## Use optimal full well mode for sensor: low full well mode.
## Step phantom vertically:
### 9 different heights.
### 5 mm vertical spacing.
### Start 20 mm below tumour, step phantom downwards until beam 20 mm above tumour.
## 8 changes in room plus initial setup; 18 total measurements (Carbon+Helium for each position).
<!-- -->
# '''High energy Helium Bragg peak scan with absorber.'''
## Switch to higher Helium energies (remote controlled by accelerator crew):
### Carbon: energy 324.26 MeV/u, intensity 8*10^7, spot size 8mm.
### Helium: energy 324.26 MeV/u, intensity 7*10^8, spot size 7mm.
## Detector measurements in both high and low full well mode.
## Build up 50 cm of PMMA in front of detector. Remove PMMA stack in 10cm steps: '''final measurement with 10cm, 5cm and 0cm absorber in place to give shoot-through calibration'''. 
##7 PMMA steps, Carbon+Helium, high+low full well mode: 28 total measurements.
## '''After first measurement with 50cm WET absorber, measure activation of PMMA and detector with full stack in place.'''
<!-- -->
# '''ADAM PETer phantom measurements.'''
## Use sensor high and low full well modes.
## Continue with higher Helium energies:
### Carbon: energy 324.26 MeV/u, intensity 8*10^7, spot size 8mm.
### Helium: energy 324.26 MeV/u, intensity 7*10^8, spot size 7mm.
## '''Setup:''' Place phantom in upright position. Fill with oil (measurements need to be fast, phantom heavily leaking oil). Place 2cm PMMA degrader in front of phantom; 15cm PMMA absorber between phantom and detector. Energy degraded down to ~300 MeV/u at phantom. Upstream face of Zarges case 28cm from isocentre.
## Align ADAM PETer in isocenter:
### Vertical shift by 5 mm.
### Start at -1cm and step upwards.
### 5 positions, 10 measurements, 4 changes in room.
## Align ADAM PETer in isocenter, rotate:
### 4 measurements (0, 2, 4, 10 degree).
### 4 measurements, 3 changes in room plus setup.
<!-- -->
# '''ADAM phantom measurements.'''
## Use optimal sensor full well mode: probably high full well.
## Continue with higher Helium energies:
### Carbon: energy 324.26 MeV/u, intensity 8*10^7, spot size 8mm.
### Helium: energy 324.26 MeV/u, intensity 7*10^8, spot size 7mm.
## Setup: Place ADAM in isocentre. 15cm PMMA absorber between phantom and detector.
## 5 measurements: Uninflated rectum, 30 Ml Air, 45 Ml Air, 60 Ml Air, 75 Ml air. Rectum ballon inserted to 15cm.
## 5 measurements, 4 changes in room.
## Align isocenter with predefined spot. Repeat previous measurements. 5 measurements, 4 changes in room.
## Repeat measurement with spot in between isocentre and predefined rectum spot (horizontal alignment isocenter, vertical 1 cm below isocenter), and spot above isocenter (horizontal alignment isocenter, vertical alignment 1cm above).
## Repeat measurement with ADAM rotation:
### Align ADAM in isocenter.
### Rotate by 2 degree steps.
## Repeat measurement with ADAM horizontal shift:
### Align ADAM in isocenter.
### Horizontal shift by 1 cm.
<!-- -->
# '''ADAM lower energy phantom measurements with AP beam (30 degree gantry rotation).'''
## Use optimal sensor full well mode: probably high full well.
## Switch to lower Helium energies:
### Carbon: energy 263.4 MeV/u, intensity 8*10^7, spot size 8mm.
### Helium: energy 263.4 MeV/u, intensity 7*10^8, spot size 7mm.
## Setup: place ADAM in isocentre in upright position.
## Repeat rectal balloon and rotation measurements.
<!-- -->
# '''Water equivalent thickness measurements of 10 x 1cm sheets:'''
## Use optimal sensor full well mode: probably high full well.
## Shoot-through calibration with 220 MeV (30cm) proton range; 5cm upstream absorber.
## Switch to monoenergetic proton beam with 10cm range.
## Insert sheets one at a time and measure range shift.
## 11 measurements, 10 setup changes in room.


== Experiment Notes ==
== Experiment Notes ==


Range calorimeter measurements with ISDI sensor + scintillator stack.
Range calorimeter measurements with ISDI sensor + photodiodes.
 
=== 12&ndash;13th April ===
 
images from sensor: User/Public/Document/EPIX/XCAP/data/20190512


images from dslr: User/Public/Document/Nikon/
Images from sensor: '''/unix/pbt/data/uclh/20210415/CMOS'''


* full sensor area
Photodiode hexadecimal data: '''/unix/pbt/data/uclh/20210415/DDC232'''
* shooting on '''BACK WINDOW'''
* Simon checking the intensity on pixels from dslr with nikon software
*camera settings
**exposure 1/2 sec
**aperture f/3
**ISO 200


{| class="wikitable"
{| class="wikitable"
!Run number
!Run number
!Full Well Mode
!Full Well Mode
!Beam Energy (MeV/u)
!Beam Energy (MeV)
!Range (estimated)
!Estimated Range (mm)
!N particles (Hz)
!Current (nA)
!Spot size (mm)
!Spot size (mm)
!ion type
!DDC232 FSR (pC)
!DSLR file name
!DDC232 Integration Time (us)
!Degrader WET (mm)
!Degrader WET (mm)
!Comments
!Comments
!Thumbnail
!PDL Comparison
!CMOS QB Fit
!PD QB Fit
!Replay Fit
!Replay Fit (Binned)
|-
|0 || high || Background || N/A || N/A || N/A || 350 || 170 || 0 || room light on || N/A || N/A || N/A || N/A || N/A
|-
|1 || high || Background || N/A || N/A || N/A || 12.5 || 170 || 0 || repeat – room light on || N/A || N/A || N/A || N/A || N/A
|-
|2 || high || Background || N/A || N/A || N/A || 12.5 || 170 || 0 || room lights off || N/A || N/A || N/A || N/A || N/A
|-
|-
 
|3 || high || 245 || 366.3 || 10 || 7.1 || 12.5 || 170 || 57.33 || shoot-through, gantry angle 270° || N/A || N/A || N/A || N/A || N/A
|N || well mode || Energy MeV/u || Range || particles || Spot FWHM || test || dslr file name DSC_ || degrader WET || comments ||  
|-
|-
|00 || low || None || - || None || None || - || - || 0 || background ||  
|4 || high || 245 || 366.3 || 20 || 7.1 || 12.5 || 170 || 57.33 || shoot-through, gantry angle 270°, CMOS sensor saturated || N/A || N/A || N/A || N/A || N/A
|-
|-
|01 || high || None || - || None || None || - || - || 0 || background ||  
|5 || high || 245 || 366.3 || 10 || 7.1 || 12.5 || 170 || 57.33 || shoot-through, gantry angle 90° || N/A || N/A || N/A || N/A || N/A
|-
|-
|not saved from Laurent || high || None || - || None || None || - || DSC_145 || 0 || intensity check ||  
|6 || high || 245 || 366.3 || 10 || 7.1 || 12.5 || 170 || 57.33 || shoot-through, gantry angle 90° || N/A || N/A || N/A || N/A || N/A
|-
|-
|none || - || - || - || - || - || - || - || 0 || '''Calibration : shoot through. no degrader''' ||  
|7 || high || 105 || || 300 || 7.1 || 12.5 || 170 || 0 || CMOS sensor saturated, BP seen || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run007.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run007.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run007.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run007.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run007%20Binned.mov]
|-
|-
|02 || high || 221.06 || - || 2*10^9 || 8.1 || p || DSC_147 || 0 || shoot through no degrader ||  
|8 || high || 105 || || 150 || 7.1 || 12.5 || 170 || 0 || CMOS sensor saturated, BP seen || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run008.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run008.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run008.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run008.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run008%20Binned.mov]
|-
|-
|03 || high || 221.06 || - || 3.2*10^9 || 11.1 || p || DSC_149 || 0 || shoot through no degrader. from now on exposure time on dslr 0.2 sec ||  
|9 || high || 70 || 40.8 || 700 || 7.1 || 12.5 || 170 || 0 || CMOS sensor saturated, BP seen || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run009.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run009.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run009.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run009.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run009%20Binned.mov]
|-
|-
|04 || high || 221.06 || - || 3.2*10^9 || 20.0 || p || DSC_150 || 0 || shoot through no degrader. ||  
|10 || high || 70  || 40.|| 350 || 7.1 || 12.5 || 170 || 0 || || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run010.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run010.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run010.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run010.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run010%20Binned.mov]
|-
|-
|05 || low || 221.06 || - || 1.2*10^9 || 20.0 || p || DSC_151 || 0 || shoot through no degrader. exposure time back to 0.5 sec ||  
|11 || high || 80 || 51.8 || 500 || 7.1 || 12.5 || 170 || 0 || Couple 1000 off saturation || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run011.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run011.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run011.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run011.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run011%20Binned.mov]
|-
|-
|06 || low || 221.06 || - || 1.2*10^9 || 11.1 || p || DSC_152 || 0 || shoot through no degrader. ||  
|12 || high || 90 || 63.9 || 400 || 7.1 || 12.5 || 170 || 0 || Saturation || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run012.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run012.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run012.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run012.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run012%20Binned.mov]
|-
|-
|07 || low || 221.06 || - || 1.2*10^9 ||  8.1 || p || DSC_153 || 0 || shoot through no degrader. ||  
|13 || high || 100 || 77.1 || 350 || 7.1 || 12.5 || 170 || 0 || || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run013.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run013.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run013.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run013.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run013%20Binned.mov]
|-
|-
|08 || high || 220.51 || - || 8*10^8 || 4.9 || He || DSC_155 || 0 || shoot through no degrader. exposure 0.1 sec  ||  
|14 || high || 110 || 91.3 || 300 || 7.1 || 12.5 || 170 || 0 || Saturation || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run014.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run014.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run014.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run014.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run014%20Binned.mov]
|-
|-
|09 || high || 220.51 || - || 8*10^8 || 8.1 || He || DSC_156 || 0 || shoot through no degrader.  ||  
|15 || high || 120 || 106.5 || 270 || 7.1 || 12.5 || 170 || 0 || BP not captured in range || N/A || N/A || N/A || N/A || N/A
|-
|-
|10 || high || 220.51 || - || 8*10^8 || 10.1 || He || DSC_157 || 0 || shoot through no degrader. ||  
|16 || high || 110 || 91.3 || 150 || 7.1 || 12.5 || 170 || 0 || || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run016.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run016.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run016.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run016.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run016%20Binned.mov]
|-
|-
|11 || high || 430.1 || - || 8*10^7 || 3.4 || C || DSC_159 || 0 || shoot through no degrader. ripple filter 3mm ||  
|17 || high || 105 ||  || 350 || 7.1 || 12.5 || 170 || 0 || Saturated || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run017.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run017.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run017.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run017.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run017%20Binned.mov]
|-
|-
|12 || high || 430.1 || - || 8*10^7 || 7.8 || C || DSC_160 || 0 || shoot through no degrader. ripple filter 3mm  ||  
|18 || high || 100 || 77.1 || 200 || 7.1 || 12.5 || 170 || 0 || Saturated || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run018.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run018.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run018.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run018.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run018%20Binned.mov]
|-
|-
|13 || high || 430.1 || - || 8*10^7 || 20 || C || DSC_161 || 0 || shoot through no degrader.ripple filter 3mm. exposure 0.5 sec ||  
|19 || high || 100 || 77.1 || 350 || 7.1 || 12.5 || 1000 || 0 || Saturated || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run019.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run019.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run019.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run019.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run019%20Binned.mov]
|-
|-
|14 || high || 430.1 || - || 8*10^7 || 3.4 || C || DSC_162 || 59 || shoot through. PMMA 5 cm ||  
|20 || high || 95 ||  || 400 || 7.1 || 12.5 || 170 || 0 || Saturated || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run020.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run020.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run020.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run020.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run020%20Binned.mov]
|-
|-
|15 || high || 430.1 || - || 8*10^7 || 7.8 || C || DSC_163 || 59 || shoot through. PMMA 5cm  ||  
|21 || high || 90 || 63.9 || 200 || 7.1 || 12.5 || 170 || 0 || || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run021.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run021.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run021.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run021.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run021%20Binned.mov]
|-
|-
|16 || high || 430.1 || - || 8*10^7 || 20 || C || DSC_164 || 59 || shoot through. PMMA 5cm  ||  
|22 || high || 90 || 63.9 || 400 || 7.1 || 12.5 || 1000 || 0 || || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run022.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run022.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run022.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run022.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run022%20Binned.mov]
|-
|-
|17 || high || 220.51|| - || 8*10^8 || 4.9 || He || DSC_165 || 59 || shoot through. PMMA 5 cm. exposure 0.1 ||  
|23 || high || 85 ||  || 500 || 7.1 || 12.5 || 170 || 0 || just missed saturation || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run023.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run023.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run023.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run023.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run023%20Binned.mov]
|-
|-
|18 || high || 220.51 || - || 8*10^8 || 8.1 || He || DSC_166 || 59 || shoot through. PMMA 5cm  ||  
|24 || high || 80 || 51.8 || 250 || 7.1 || 12.5 || 170 || 0 || || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run024.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run024.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run024.pdf] ||[http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run024.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run024%20Binned.mov]
|-
|-
|19 || high || 220.51 || - || 8*10^8 || 10.1 || He || DSC_167 || 59 || shoot through. PMMA 5cm  ||  
|25 || high || 80 || 51.8 || 500 || 7.1 || 12.5 || 1000 || 0 || || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run025.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run025.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run025.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run025.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run025%20Binned.mov]
|-
|-
|20 || high || 221.06 || - || 3.2*10^9 || 8.1 || p || DSC_168 || 59 || shoot through. PMMA 5 cm. exposure 0.2 ||  
|26 || high || 75 ||  || 600 || 7.1 || 12.5 || 170 || 0 || || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run026.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run026.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run026.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run026.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run026%20Binned.mov]
|-
|-
|21 || high || 221.06 || - || 3.2*10^9 || 11.1 || p || DSC_169 || 59 || shoot through. PMMA 5cm  ||  
|27 || high || 70 || 40.8 || 350 || 7.1 || 12.5 || 170 || 0 || || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run027.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run027.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run027.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run027.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run027%20Binned.mov]
|-
|-
|22 || high || 221.06 || - || 3.2*10^9 || 20 || p || DSC_170 || 59 || shoot through. PMMA 5cm  ||  
|28 || high || 70 || 40.8 || 700 || 7.1 || 12.5 || 1000 || 0 || || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/CMOS_PD_PDL/Run028.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/CMOS/20210415/Fits/Run028.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/images/DDC232/20210415/fits/Run028.pdf] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run028.mov] || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run028%20Binned.mov]
|-
|-
|23 || high || 430.1 || - || 8*10^7 || 3.4 || C || DSC_172 || 0 || shoot through. no absorber. exposure 0.1 ||  
|29 || N/A || Treatment Plan || N/A || N/A || N/A || 12.5 || 170 || 0 || clinical mode from now - 6Gy scanned box field || N/A || N/A || N/A || N/A || N/A
|-
|-
|24 || high || 430.1 || - || 8*10^7 || 7.8 || C || DSC_173 || 0 || shoot through. no absorber ||  
|30 || N/A || Treatment Plan || N/A || N/A || N/A || 12.5 || 170 || 0 || repeat || N/A || N/A || N/A || [http://www.hep.ucl.ac.uk/pbt/wikiData/movies/DDC232/210415/Run030%20Replay%20Calibrated%20Half-Speed.mov]|| N/A
|-
|-
|25 || high || 430.1 || - || 8*10^7 || 20 || C || DSC_175 || 0 || shoot through. no absorber. exposure 0.2 sec  ||  
|31 || N/A || Treatment Plan || N/A || N/A || N/A || 50 || 170 || 0 || || N/A || N/A || N/A || N/A || N/A
|-
|-
|}

Latest revision as of 12:01, 8 May 2023

Afternoon test at UCLH with photodiodes and CMOS sesnor.

Experiment Equipment

Item Notes
Network Hub Set in control room to take output from experimental room ethernet connection. Control laptops connected via 5GHz WiFi.
Control Laptop x3 2x remote control for sensor and FPGA, 1 for notes/web GUI.
Ethernet Cable x 3 To connect DAQ laptops to network in the experimental room, network hub to network in control room.
Portable Enclosure Modified Big Zarges Waterproof Wheeled Equipment Case.

Features mount for scintillator stack, front and back openings for beam, patch panel with ports for SHV, BNC, SMA, USB, Camera Link cables.
Hole for sensor powering.
Mylar windows with Al support mounted to internal face: approximately light-tight.

Scintillator stack X x 10 mm, Y x 10 mm and Z x (approx) 2.8 mm sheets in ascending order from the front/beam of the end of the scintillator. Sheet numbers from front to end (in beam direction gantry angle 90°): 25, 31, 50, 52, 34, 44, 53, 47, 33, 54, 46, 42, 55, 45, 38, 36, 43, 32, 48, 41, 8, 39, 40, 37, 35, 49, 9, 2, 1, 51, 11, 4, 3, 23, 6. Full stack thickness measurements (in mm): 98.10, 98.08, 97.93, 98.49.
ISDI CMOS sensor sensor pixel dimension: 1030 x 1536. NO optical grease between scintillator and sensor. Connected to DAQ PC via Camera Link cable.
Nikon D7500 Connected to DAQ Laptop via USB cable.
DAQ laptop x2 Control sensor and photodiode acquisition.
Nexys Video FPGA development board. For interfacing between DDC232 and PC.
Rev. 2 Texas Instruments DDC232 custom circuit board (x2) Each housing 16x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets. No optical gel and aligned by eye to scintillator sheets. Photodiodes do not cover sheets 25, 23 and 6.
Gloves For handling scintillator

Schematic

UCLH_Setup.png

ISDI Sensor Instructions

General Sensor Configuration

  • Base Configuration is 14 x 1 Bit
  • Resolution is 1030 x 1536 pixels
  • For selecting pixel rows in centre: Use horizontal offset (ignore DVAL to activate field) and horizontal resolution

Increase number of frame buffers

  • To increase the number of frame buffers to the allowed maximum, start XCAP as administrator.
  • Select Capture/Sequence Capture/Video to Frame Buffers/Driver Assistant
  • Tick box in left field, then set number of frame buffers here.

Serial Terminal instructions for ISDI sensor

  • Find serial terminal in PIXCI(R) tab of sensor control software.
  • In Controls/Options: Make sure "Send string with CR" is selected. Should be default.
  • In Controls/Setup: Select Serial Port: Enabled. Baud rate has to be 115200
  • Set low full well mode using W000300000
  • Set high full well mode using W000300004
  • Read full well mode using R0003

Continuous Field buffer setting

  • This setting makes sure that the buffer is filled with a new frame as they come in. If this option is not selected the sensor and software will not be synchronised and we *will record a lot of pitch black images.
  • Select Capture/Live Options//Live Mode/Live Video: Continuous All buffer capture

DDC232 Instructions

  • Bash script prog.sh executes FTDI.cpp and livePlot.cpp. Execute with . prog.sh 2 350 170 0 50 true Data/capture.txt in the directory Photodiodes.
    • First argument is the number of DDC232s.
    • Second argument is the full-scale range (FSR) in pC. Choices are: 12.5, 50, 100, 150, 200, 250, 300, 350.
    • Third argument is the integration time in us. Range is 170-1455. Set in multiples of 5 us.
    • Fourth argument is the number of measurements to acquire. Typically acquire 5700 measurements per second. Set to 0 for infinite loop.
    • Fifth argument is the rate in Hz to update the file average.csv for ROOT/web GUI plotting.
    • Sixth argument is a boolean to choose whether to display the ROOT plot.
    • Seventh argument is the filename and destination for the output hexadecimal data.
  • After acquisition, close ROOT by choosing File->Quit in the ROOT window and then stopping DAQ (if in infinite loop) with CTRL-C.

Experiment Plan

  • Shoot-through measurements (back & front) for light calibration and level setting.
  • Range measurements between 70-120 MeV with 1mm steps with CMOS sensor and photodiodes.
  • High intensity range measurements between 70-120 MeV with 1mm steps with photodiodes at optimal light level.
  • Live measurement of 3cm x 3cm x 3cm treatment plan.

Experiment Notes

Range calorimeter measurements with ISDI sensor + photodiodes.

Images from sensor: /unix/pbt/data/uclh/20210415/CMOS

Photodiode hexadecimal data: /unix/pbt/data/uclh/20210415/DDC232

Run number Full Well Mode Beam Energy (MeV) Estimated Range (mm) Current (nA) Spot size (mm) DDC232 FSR (pC) DDC232 Integration Time (us) Degrader WET (mm) Comments PDL Comparison CMOS QB Fit PD QB Fit Replay Fit Replay Fit (Binned)
0 high Background N/A N/A N/A 350 170 0 room light on N/A N/A N/A N/A N/A
1 high Background N/A N/A N/A 12.5 170 0 repeat – room light on N/A N/A N/A N/A N/A
2 high Background N/A N/A N/A 12.5 170 0 room lights off N/A N/A N/A N/A N/A
3 high 245 366.3 10 7.1 12.5 170 57.33 shoot-through, gantry angle 270° N/A N/A N/A N/A N/A
4 high 245 366.3 20 7.1 12.5 170 57.33 shoot-through, gantry angle 270°, CMOS sensor saturated N/A N/A N/A N/A N/A
5 high 245 366.3 10 7.1 12.5 170 57.33 shoot-through, gantry angle 90° N/A N/A N/A N/A N/A
6 high 245 366.3 10 7.1 12.5 170 57.33 shoot-through, gantry angle 90° N/A N/A N/A N/A N/A
7 high 105 300 7.1 12.5 170 0 CMOS sensor saturated, BP seen [1] [2] [3] [4] [5]
8 high 105 150 7.1 12.5 170 0 CMOS sensor saturated, BP seen [6] [7] [8] [9] [10]
9 high 70 40.8 700 7.1 12.5 170 0 CMOS sensor saturated, BP seen [11] [12] [13] [14] [15]
10 high 70 40.8 350 7.1 12.5 170 0 [16] [17] [18] [19] [20]
11 high 80 51.8 500 7.1 12.5 170 0 Couple 1000 off saturation [21] [22] [23] [24] [25]
12 high 90 63.9 400 7.1 12.5 170 0 Saturation [26] [27] [28] [29] [30]
13 high 100 77.1 350 7.1 12.5 170 0 [31] [32] [33] [34] [35]
14 high 110 91.3 300 7.1 12.5 170 0 Saturation [36] [37] [38] [39] [40]
15 high 120 106.5 270 7.1 12.5 170 0 BP not captured in range N/A N/A N/A N/A N/A
16 high 110 91.3 150 7.1 12.5 170 0 [41] [42] [43] [44] [45]
17 high 105 350 7.1 12.5 170 0 Saturated [46] [47] [48] [49] [50]
18 high 100 77.1 200 7.1 12.5 170 0 Saturated [51] [52] [53] [54] [55]
19 high 100 77.1 350 7.1 12.5 1000 0 Saturated [56] [57] [58] [59] [60]
20 high 95 400 7.1 12.5 170 0 Saturated [61] [62] [63] [64] [65]
21 high 90 63.9 200 7.1 12.5 170 0 [66] [67] [68] [69] [70]
22 high 90 63.9 400 7.1 12.5 1000 0 [71] [72] [73] [74] [75]
23 high 85 500 7.1 12.5 170 0 just missed saturation [76] [77] [78] [79] [80]
24 high 80 51.8 250 7.1 12.5 170 0 [81] [82] [83] [84] [85]
25 high 80 51.8 500 7.1 12.5 1000 0 [86] [87] [88] [89] [90]
26 high 75 600 7.1 12.5 170 0 [91] [92] [93] [94] [95]
27 high 70 40.8 350 7.1 12.5 170 0 [96] [97] [98] [99] [100]
28 high 70 40.8 700 7.1 12.5 1000 0 [101] [102] [103] [104] [105]
29 N/A Treatment Plan N/A N/A N/A 12.5 170 0 clinical mode from now - 6Gy scanned box field N/A N/A N/A N/A N/A
30 N/A Treatment Plan N/A N/A N/A 12.5 170 0 repeat N/A N/A N/A [106] N/A
31 N/A Treatment Plan N/A N/A N/A 50 170 0 N/A N/A N/A N/A N/A