Line 101: Line 101:
* Run0xy_fitted_25Hz.txt: data averaged at 25Hz and fitted, with range and energy values extracted. This can be visualised in the REPLAY.html tab of the GUI.
* Run0xy_fitted_25Hz.txt: data averaged at 25Hz and fitted, with range and energy values extracted. This can be visualised in the REPLAY.html tab of the GUI.
   
   
* 50mm solid water absorber installed in front of QuARC for shoot-through calibration measurements?
* 50mm solid water absorber installed in front of QuARC for shoot-through calibration measurements
 
* In between Run 38 and 39, measurements taken with IBA Zebra with no QuARC and just 50 mm solid water than just Zebra (no absorber).


{| class="wikitable"
{| class="wikitable"
Line 114: Line 116:
!Estimated Range  (mm)
!Estimated Range  (mm)
!Comments
!Comments
!Photodiode QB Fit
|-
|-
! colspan="12" | Saturday 26/04/2025
! colspan="12" | Saturday 26/04/2025
Line 120: Line 121:
! colspan="12" | Machine QA Mode
! colspan="12" | Machine QA Mode
|-
|-
|0  || No stack - just peli case || 226MeV (shoot through front) || 8 nA (?) || 30000 || 12.5 || 170 (constant) || - || - || Without 50 mm solid water absorber || -  
|0  || No stack - just peli case || 226MeV (shoot through front) || 8 nA (?) || 30000 || 12.5 || 170 (constant) || - || - || Without 50 mm solid water absorber
|-
|1  || Stack #1 || 226MeV (shoot through front) || 8 nA (?) || 30000 || 12.5 || 170 (constant) || - || - || Without 50 mm solid water absorber and beam on time too short so only reading background for last measurements
|-
|2  || Stack #1 || 245MeV (shoot through front) || 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||With 50 mm solid water absorber and beam mode changed to continuous i.e. beam on before starting and after ending measurements) at clinical current (8 nA)
|-
|3  || Stack #1 || 245MeV (shoot through front) || 8 nA || 30000 || 12.5 || 170 (constant) || - || - || repeat of run002 - zebra measurement: 16.76 cm
|-
|4  || Stack #1 || background (back pos.) || || 30000 || 12.5 || 170 (constant) || - || - || -
|-
|5  || Stack #1 || 245MeV (shoot through back) || 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  zebra measurement: 16.76 cm
|-
|6  || Stack #2 || 245MeV (shoot through front) || 8 nA || 30000 || 12.5 || 170 (constant) || - || - || note dead photodiode on centre of left board. zebra measurement: 16.68 cm 
|-
|7  || Stack #2 || background (back)||  nA || 30000 || 12.5 || 170 (constant) || - || - || -
|-
|8  || Stack #2 || shoot through back|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - || -
|-
|9  || Stack #3 || background front ||  nA || 30000 || 12.5 || 170 (constant) || - || - || -
|-
|10  || Stack #3 || shoot through front|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - || zebra measurement: cm
|-
|11  || Stack #3 || background back||  || 30000 || 12.5 || 170 (constant) || - || - || -
|-
|12  || Stack #3 || shoot through back|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  zebra measurement: cm
|-
|13  || Stack #4 || background front||  nA || 30000 || 12.5 || 170 (constant) || - || - || -
|-
|14  || Stack #4 || shoot through front || 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  zebra measurement: cm
|-
|15  || Stack #4 || background back ||  nA || 30000 || 12.5 || 170 (constant) || - || - || -  
|-
|-
|1 || Stack #1 || 226MeV (shoot through front) || 8 nA (?) || 30000 || 12.5 || 170 (constant) || - || - || Without 50 mm solid water absorber and beam on time too short so only reading background for last measurmements || -
|16 || Stack #4 || shoot through back || 8 nA || 30000 || 12.5 || 170 (constant) || - || - || zebra measurement: cm
|-
|-
|2 || Stack #1 || 245MeV (shoot through front) || 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||With 50 mm solid water absorber and beam mode changed to continuous 9i.e. beam on before starting and after ending measurements) at clinical current (8 nA) || -
|17 || Stack #1 -2  || background || || 30000 || 12.5 || 170 (constant) || - || - || noise on cabling 
|-
|-
|3 || Stack #1 || 245MeV (shoot through front) || 8 nA || 30000 || 12.5 || 170 (constant) || - || - || repeat of run002 - zebra measurement: 16.76 cm || -  
|18 || Stack #1-4 || background || nA || 30000 || 12.5 || 170 (constant) || - || - || -  
|-
|-
|4 || Stack #1 || background (back pos.) || || 30000 || 12.5 || 170 (constant) || - || - ||   || -
|19 || Stack #1-4 || 226 MeV (max)|| 140 nA || 30000 || 12.5 || 170 (constant) || - || - || just 4 boards (so one sided) but acquiring with requesting 8
|-
|-
|5 || Stack #1 || 245MeV (shoot through back) || 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  zebra measurement: 16.76 cm || -
|20 || Stack #1-4 || 226 MeV (max)|| 140 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8. attempting applying one sided calibration
|-
|-
|6 || Stack #2 || 245MeV (shoot through front) || 8 nA || 30000 || 12.5 || 170 (constant) || - || - || note dead photodiode on centre of left board zebra measurement: 16.68 cm  || -
|21 || Stack #1-4 || 226 MeV (max)|| 140 nA || 30000 || 12.5 || 170 (constant) || - || - || just 4 boards (so one sided) but acquiring with requesting 8. attempting applying one sided calibration
|-
|-
|7 || Stack #2 || background (back)|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||   || -  
|22 || Stack #1-4 || background|| 140 nA || 30000 || 12.5 || 170 (constant) || - || - || just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|8 || Stack #2 || shoot through back|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||   || -  
|23 || Stack #1-4 || 200 MeV (?) or 226 MeV|| 140 nA || 30000 || 12.5 || 170 (constant) || - || - || just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|8 || Stack #2 || background front || 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  || -  
|24 || Stack #1-4 || 180 MeV (?) or 200 MeV || 140 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|10 || Stack #3 || shoot through front|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  zebra measurement: cm || -  
|25 || Stack #1-4 || 180 MeV|| 140 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|11 || Stack #3 || background back|| || 30000 || 12.5 || 170 (constant) || - || - ||  || -  
|26 || Stack #1-4 || 160 MeV|| 140 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|12 || Stack #3 || shoot through back|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  zebra measurement: cm || -  
|27 || Stack #1-4 || 160 MeV|| 140 nA || 30000 || 12.5 - 50 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration beam lost part way at end and FSR wasn't fully updated at start?
|-
|-
|13 || Stack #4 || background front| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||   || -  
|28 || Stack #1-4 || 160 MeV|| 140 nA || 30000 || 50 || 170 (constant) || - || - || just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration|
|-
|29  || Stack #1-4 || 140 MeV|| 140 nA || 30000 || 50 || 170 (constant) || - || - || just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration. Wanting to drop current as starting to see saturation on boards - was not aware they were operating last around 140 nA for pristine Bragg peaks - had assumed 8 nA (clinical current)
|-
|-
|14 || Stack #4 || shoot through front || 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  zebra measurement: cm || -  
|30 || Stack #1-4 || 120 MeV|| 40 nA || 30000 || 50 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|15 || Stack #4 || background back|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  || -  
|31 || Stack #1-4 || 120 MeV|| 40 nA || 30000 || 50 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|16 || Stack #4 || shoot through back|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  zebra measurement: cm || -  
|32 || Stack #1-4 || background|| || 30000 || 50 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|17 || Stack #1 -|| background|| || 30000 || 12.5 || 170 (constant) || - || - ||  noise on cabling || -  
|33 || Stack #1-4 || 100 MeV || 40 nA || 30000 || 50 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|18 || Stack #1-4 || background|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||   || -  
|34 || Stack #1-4 || 80 MeV|| 40 nA || 30000 || 50 || 170 (constant) || - || - || just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|19 || Stack #1-4 || 226 MeV (max)|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 || -  
|35 || Stack #1-4 || 80 MeV|| 40 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|20 || Stack #1-4 || 226 MeV (max)|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - attempting applying one sided calibration || -
|36 || Stack #1-4 || background|| || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|-
|-
|21 || Stack #1-4 || 226 MeV (max)|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - attempting applying one sided calibration|| -
|37 || Stack #1-4 || background||  || 30000 || 50 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
|22  || Stack #1-4 || background|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration|| -
|-
|-
|23 || Stack #1-4 || 200 MeV|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration|| -  
|38 || Stack #1-4 || Single spot SOBP across the entire range of energies that can be obtained from the nozzle (). 5 Gy|| ? nA || 1 min ( stopped at 40%) || 50 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration - total plan time was about 34s, started before beam stopped at 40% -  appears display rate of GUI was too slow - raw data saved fine. Note that Vlad said the TPS for this plan sees the single spot SOBP as 3 fields and superimposes them to create SOBP.
|-
|-
|24 || Stack #1-4 || 180 MeV (?)|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - || just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration|| -  
|39 || Stack #1-2 (dual-sided with 4 boards) || background || || 30000 || 12.5 || 170 (constant) || - || - || moving to two modules with 4 boards. Braided USB-C to USB-C cable used to connect the two boards on stack 2 via the back of the stack.
|-
|-
|25 || Stack #1-4 || 180 MeV|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration|| -  
|40 || Stack #1-2 (dual-sided with 4 boards) || 160-120 MeV SOBP || nA || 1 min ( ) || 50 || 170 (constant) || - || -  
|-
|-
|26 || Stack #1-4 || 160 MeV|| 8 nA || 30000 || 12.5 || 170 (constant) || - || - ||  just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration|| -  
|41 || Stack #1-2 (dual-sided with 4 boards) || background || || 30000 || 50 || 170 (constant) || - || -
|-
|-
|27 || Stack #1-4 || 160 MeV|| 8 nA || 30000 || 12.5 - 50 || 170 (constant) || - || - || just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration beam lost part way at end and FSR wasn't fully updated at start? || -
|42 || Stack #1-2 (dual-sided with 4 boards) || 160 MeV pristine Bragg beak|| 40 nA || 30000 || 50 || 170 (constant) || - || - || 10s beam on time
|-
|-
|28 || Stack #1-4 || 160 MeV|| 8 nA || 30000 || 50 || 170 (constant) || - || - || just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration|| -
|43 || Stack #1-2 (dual-sided with 4 boards) || 20 Gy 5x5 box field || nA || 1 min ( ) || 50 || 170 (constant) || - || - || around 300s beam on time, set QuARC acquisition time to 900s and stopped around 35 %

Latest revision as of 14:37, 29 April 2025

Experiment Equipment

Item Notes
Portable Enclosure Modified Small Peli Case Wheeled. Features mount for scintillator stacks, front and back openings for beam, patch panel with ports. Mylar windows with 3D printed alignment plates: approximately light-tight.
Rev. E Scintillator stack 1 (with Rev. E boards no. 1 (L) and 4 (R)) 32 X x 105 mm, Y x 105 mm and Z x (approx) 3 mm sheets (mix of opaque and machined sheets).
Rev. E Scintillator stack 2 (with Rev. E boards no. 3 (L) and 9 (R)) 32 X x 105 mm, Y x 105 mm and Z x (approx) 3 mm sheets (just machined sheets with the closest to ideal thicknesses ~3mm).
Rev. E Scintillator stack 3 (with Rev. E boards no. 5 (L) and 6 (R)) 32 X x 105 mm, Y x 105 mm and Z x (approx) 3 mm sheets (mix of opaque and machined sheets).
Rev. E Scintillator stack 4 (spare stack not installed) 32 X x 105 mm, Y x 105 mm and Z x (approx) 3 mm sheets (mix of opaque, machined, and injection moulded sheets - stacked in pairs in alternating directions).
Rev. E Scintillator stack 5 (with Rev. E boards no. 7 (L) and 10 (R)) 32 X x 105 mm, Y x 105 mm and Z x (approx) 3 mm sheets mix of opaque, machined, and injection moulded sheets - stacked in pairs in alternating directions).
Rev. E Texas Instruments DDC232 custom circuit board (x9 - 8 installed in modules 1,2,3,5 and one spare) Housing 32x Hamamatsu S12915-16R photodiodes, coupled directly to scintillator sheets. Aligned by eye to scintillator sheets.
USB104 FPGA development board x 2 (1 main and 1 spare) For interfacing between DDC232 and pi.
Raspberry pi5 x 2 (1 main and 1 spare) Control data acquisition + visualise GUI.
Power Strip: To connect FPGA and daughter board.
DAQ laptop x1 Back up option for data acquisition.
Control Laptop x2 1 for remote control (Sonia's MacBook Pro), 1 for notes/web GUI (Joe's MacBook Pro).
D-LINK Network Hub Set in experimental room to connect pi5, DAQ laptop + NPL devices
GL-Inet Router Set in control room to take output from experimental room ethernet connection. Control laptops connected via ethernet or 5GHz WiFi.
Ethernet Cable x4 (?) To connect pi5 and DAQ laptop to network in the experimental room, hub to wall and in control room router to wall, and router to laptop.
MacBook Adapters x2: one for pi5 DAQ laptop one for DAQ laptop
Spare Cables: Usb-C flexible cables, USB-C corded cables (and right-angles angled connector), USB-C to USB-A adapter, USB-C to USB-B adaptor, USB-A to USB-B cable, right-angles network cables. power cables
Gloves For handling scintillator

Schematic

Experiment Plan

Saturday 26th April


1) Shoot-through of each scintillator stack at 250 MeV to check prior calibration. 50mm absorber (?) -> scint stack -> water dump/IBA Zebra or Giraffe for simultaneous WET measurement. Both directions for stack so will rotate midway through.

2) Background measurement of scintillator stacks 1 - 5 (exc. 4 which is spare module)

-- UCL: Add calibration files to appropiate file for GUI

3) Rotate gantry to front of case. Full system check at 245 MeV with beam on axis.

4) Single spot SOBP across the entire range of energies that can be obtained from the nozzle. Beam on axis.

5) Single spot SOBP from 165 MeV to 120 MeV. Beam on axis.

3) 5 x 5 x 5cm box field, on axis from 12cm to 17cm deep.

4) Single layer with 5 x 5 grid of spots, 1cm spacing, with single spot dwell time of 500ms–1s


30,000 measurements ~ 5 secs for pristine Bragg peaks/shoot throughs.

1-2 mins per SOBP.

Experiment Notes

Original data will be saved in pi5: /home/pi/GUI/v3.01b/

A copy ot the hexadecimal files will be saved to: /unix/pbt/data/

For each run ... files will be generated:

  • Run0xy.txt: raw hexadecimal data
  • RunOxy_calibrated.txt: calibrated data with background subtraction. An average measurement is at the bottom with a run performed to it. Visualisation with acquisition on DAQ.html
  • Run0xy_fitted_25Hz.txt: data averaged at 25Hz and fitted, with range and energy values extracted. This can be visualised in the REPLAY.html tab of the GUI.
  • 50mm solid water absorber installed in front of QuARC for shoot-through calibration measurements
  • In between Run 38 and 39, measurements taken with IBA Zebra with no QuARC and just 50 mm solid water than just Zebra (no absorber).
Run number Stack No. Beam Energy (MeV) Current (nA) Measurements DDC232 FSR (pC) DDC232 Integration Time (us) Spot size (mm, FWHM) Estimated Range (mm) Comments
Saturday 26/04/2025
Machine QA Mode
0 No stack - just peli case 226MeV (shoot through front) 8 nA (?) 30000 12.5 170 (constant) - - Without 50 mm solid water absorber
1 Stack #1 226MeV (shoot through front) 8 nA (?) 30000 12.5 170 (constant) - - Without 50 mm solid water absorber and beam on time too short so only reading background for last measurements
2 Stack #1 245MeV (shoot through front) 8 nA 30000 12.5 170 (constant) - - With 50 mm solid water absorber and beam mode changed to continuous i.e. beam on before starting and after ending measurements) at clinical current (8 nA)
3 Stack #1 245MeV (shoot through front) 8 nA 30000 12.5 170 (constant) - - repeat of run002 - zebra measurement: 16.76 cm
4 Stack #1 background (back pos.) 30000 12.5 170 (constant) - - -
5 Stack #1 245MeV (shoot through back) 8 nA 30000 12.5 170 (constant) - - zebra measurement: 16.76 cm
6 Stack #2 245MeV (shoot through front) 8 nA 30000 12.5 170 (constant) - - note dead photodiode on centre of left board. zebra measurement: 16.68 cm
7 Stack #2 background (back) nA 30000 12.5 170 (constant) - - -
8 Stack #2 shoot through back 8 nA 30000 12.5 170 (constant) - - -
9 Stack #3 background front nA 30000 12.5 170 (constant) - - -
10 Stack #3 shoot through front 8 nA 30000 12.5 170 (constant) - - zebra measurement: cm
11 Stack #3 background back 30000 12.5 170 (constant) - - -
12 Stack #3 shoot through back 8 nA 30000 12.5 170 (constant) - - zebra measurement: cm
13 Stack #4 background front nA 30000 12.5 170 (constant) - - -
14 Stack #4 shoot through front 8 nA 30000 12.5 170 (constant) - - zebra measurement: cm
15 Stack #4 background back nA 30000 12.5 170 (constant) - - -
16 Stack #4 shoot through back 8 nA 30000 12.5 170 (constant) - - zebra measurement: cm
17 Stack #1 -2 background 30000 12.5 170 (constant) - - noise on cabling
18 Stack #1-4 background nA 30000 12.5 170 (constant) - - -
19 Stack #1-4 226 MeV (max) 140 nA 30000 12.5 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8
20 Stack #1-4 226 MeV (max) 140 nA 30000 12.5 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8. attempting applying one sided calibration
21 Stack #1-4 226 MeV (max) 140 nA 30000 12.5 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8. attempting applying one sided calibration
22 Stack #1-4 background 140 nA 30000 12.5 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
23 Stack #1-4 200 MeV (?) or 226 MeV 140 nA 30000 12.5 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
24 Stack #1-4 180 MeV (?) or 200 MeV 140 nA 30000 12.5 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
25 Stack #1-4 180 MeV 140 nA 30000 12.5 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
26 Stack #1-4 160 MeV 140 nA 30000 12.5 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
27 Stack #1-4 160 MeV 140 nA 30000 12.5 - 50 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration beam lost part way at end and FSR wasn't fully updated at start?
28 Stack #1-4 160 MeV 140 nA 30000 50 170 (constant) - -
29 Stack #1-4 140 MeV 140 nA 30000 50 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration. Wanting to drop current as starting to see saturation on boards - was not aware they were operating last around 140 nA for pristine Bragg peaks - had assumed 8 nA (clinical current)
30 Stack #1-4 120 MeV 40 nA 30000 50 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
31 Stack #1-4 120 MeV 40 nA 30000 50 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
32 Stack #1-4 background 30000 50 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
33 Stack #1-4 100 MeV 40 nA 30000 50 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
34 Stack #1-4 80 MeV 40 nA 30000 50 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
35 Stack #1-4 80 MeV 40 nA 30000 12.5 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
36 Stack #1-4 background 30000 12.5 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
37 Stack #1-4 background 30000 50 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration
38 Stack #1-4 Single spot SOBP across the entire range of energies that can be obtained from the nozzle (). 5 Gy ? nA 1 min ( stopped at 40%) 50 170 (constant) - - just 4 boards (so one sided) but acquiring with requesting 8 - sticking with displaying uncalibrated data and will post-process calibration - total plan time was about 34s, started before beam stopped at 40% - appears display rate of GUI was too slow - raw data saved fine. Note that Vlad said the TPS for this plan sees the single spot SOBP as 3 fields and superimposes them to create SOBP.
39 Stack #1-2 (dual-sided with 4 boards) background 30000 12.5 170 (constant) - - moving to two modules with 4 boards. Braided USB-C to USB-C cable used to connect the two boards on stack 2 via the back of the stack.
40 Stack #1-2 (dual-sided with 4 boards) 160-120 MeV SOBP nA 1 min ( ) 50 170 (constant) - -
41 Stack #1-2 (dual-sided with 4 boards) background 30000 50 170 (constant) - -
42 Stack #1-2 (dual-sided with 4 boards) 160 MeV pristine Bragg beak 40 nA 30000 50 170 (constant) - - 10s beam on time
43 Stack #1-2 (dual-sided with 4 boards) 20 Gy 5x5 box field nA 1 min ( ) 50 170 (constant) - - around 300s beam on time, set QuARC acquisition time to 900s and stopped around 35 %