| {| class="wikitable" | |||
| !Run number  | |||
| !Detector | |||
| !Beam Energy (MeV) | |||
| !Estimated Range (mm) | |||
| !Current (nA) | |||
| !Spot size (mm, FWHM) | |||
| !DDC232 FSR (pC) | |||
| !DDC232 Integration Time (us) | |||
| !Degrader WET (mm) | |||
| !Comments | |||
| !Photodiode QB Fit | |||
| !Replay Fit | |||
| |- | |||
| |3 || Stack 1 || 245 MeV|| || max || || min FSR that does not saturate the signal || 170 || ||  Shoot through front || N/A ||  | |||
| |- | |||
| |4 || Stack 2 || Background || || - || N/A || 350 || 170 ||  ||  || N/A ||  | |||
| |- | |||
| |5 || Stack 2 || 245 MeV || || 800 || 4.0 ||  350 || 170 || 50 || Shoot through front   || N/A ||  | |||
| |- | |||
| |6 || Stack 3 || Background || || - || N/A ||  350 || 170 ||  ||   || N/A ||  | |||
| |- | |||
| |7 || Stack 3 || 245 MeV || || 800 || 4.0 ||  350 || 170 || 50 || Shoot through front  || N/A ||  | |||
| |- | |||
| |8 || Stack  4|| Background || || - || N/A || 350 || 170 ||  ||  || N/A ||  | |||
| |- | |||
| |9 || Stack 4 || 245 MeV || || 800 || 4.0 ||  350 || 170 || 50 || Shoot through front   || N/A ||  | |||
| |- | |||
| |10 || Stack 1 || Background || || - || N/A ||  350 || 170 ||  ||   || N/A ||  | |||
| |- | |||
| |11 || Stack 1|| 245 MeV || || 800 || 4.0 || 350 || 170 || 50 ||  Shoot through back|| N/A ||  | |||
| |- | |||
| |12 || Stack  2|| 245 MeV || || 800 || 4.0 ||  350 || 170 || 50 || Shoot through back   || N/A ||  | |||
| |- | |||
| |13 || Stack 3 ||  245 MeV || || 800 || 4.0 ||  350 || 170 || 50 || Shoot through back   || N/A || | |||
| |- | |||
| |14 || Stack 4 ||  245 MeV || || 800 || 4.0 ||  350 || 170 || 50 || Shoot through back   || N/A || | |||
| |- | |||
| |} | |||
Revision as of 10:28, 4 February 2025
Experiment Equipment
| Item | Notes | 
|---|---|
| Network Hub | Set in control room to take output from experimental room ethernet connection. Control laptops connected via 5GHz WiFi. | 
| Control Laptop x2 | 1 for remote control of FPGA, 1 for notes/web GUI. | 
| Ethernet Cable x 2 | To connect DAQ laptop to network in the experimental room, control laptop to network in control room. | 
| Portable Enclosure | |
| Scintillator stacks | 4 Modules (to cover the full clinical range, from 60 to 250 MeV) | 
| Scintillating fibers module | 1 2D Module | 
| DAQ laptop x1 | Control photodiode acquisition. | 
| Nexys Video FPGA development board. | For interfacing between DDC232 and PC. | 
Schematic
Experiment Plan
- Set up mains cabling and networking in treatment and control rooms.
- no beam
 
- Install Peli case on treatment couch.
- no beam
 
- Background
- no beam
 
- Calibration shoot-through measurements of 4 stack modules: either rotating case or modules for back shoot-through.
- 8 beam deliveries (2 beam deliveries (back and forth) per each module):
 
| Run number | Detector config | Beam Energy (MeV) | Beam Current (nA) | Physical thickness (mm) | D80 measured (mm) | WET (mm) | RSP | Comments | 
|---|---|---|---|---|---|---|---|---|
| 33 | Stack 1 front | 245 | 18.8 (18.7) | 95.95 | 198.69 | 98.24 | 1.024 | |
| 34 | Stack 1 back | 245 | 18.8 (18.7) | 95.95 | 198.74 | 98.19 | 1.023 | |
| 36 | Stack 2 front | 245 | 18.8 (18.7) | 96.81 | 198.51 | 98.42 | 1.016 | |
| 37 | Stack 3 front | 245 | 18.8 (18.7) | 95.48 | 199.78 | 97.15 | 1.017 | |
| 40 | Stack 4 front | 245 | 18.8 (18.7) | 94.07 | 202.26 | 94.67 | 1.006 | |
| 41 | Empty Box | 245 | 18.8 (18.7) | 0.012 | 296.93 | 0 | 0 | |
| 42 | No Box | 245 | 18.8 (18.7) | 0 | 296.90 | 0 | 0 | 
- Pristine Bragg peak measurements at clinical current for full clinical range.
- FLASH measurements at 245 MeV.
Calibration
| Run number | Detector | Beam Energy (MeV) | Estimated Range (mm) | Current (nA) | Spot size (mm, FWHM) | DDC232 FSR (pC) | DDC232 Integration Time (us) | Degrader WET (mm) | Comments | Photodiode QB Fit | Replay Fit | 
|---|---|---|---|---|---|---|---|---|---|---|---|
| 3 | Stack 1 | 245 MeV | max | min FSR that does not saturate the signal | 170 | Shoot through front | N/A | ||||
| 4 | Stack 2 | Background | - | N/A | 350 | 170 | N/A | ||||
| 5 | Stack 2 | 245 MeV | 800 | 4.0 | 350 | 170 | 50 | Shoot through front | N/A | ||
| 6 | Stack 3 | Background | - | N/A | 350 | 170 | N/A | ||||
| 7 | Stack 3 | 245 MeV | 800 | 4.0 | 350 | 170 | 50 | Shoot through front | N/A | ||
| 8 | Stack 4 | Background | - | N/A | 350 | 170 | N/A | ||||
| 9 | Stack 4 | 245 MeV | 800 | 4.0 | 350 | 170 | 50 | Shoot through front | N/A | ||
| 10 | Stack 1 | Background | - | N/A | 350 | 170 | N/A | ||||
| 11 | Stack 1 | 245 MeV | 800 | 4.0 | 350 | 170 | 50 | Shoot through back | N/A | ||
| 12 | Stack 2 | 245 MeV | 800 | 4.0 | 350 | 170 | 50 | Shoot through back | N/A | ||
| 13 | Stack 3 | 245 MeV | 800 | 4.0 | 350 | 170 | 50 | Shoot through back | N/A | ||
| 14 | Stack 4 | 245 MeV | 800 | 4.0 | 350 | 170 | 50 | Shoot through back | N/A | 