Line 67: Line 67:




| 5 || f/7.1 || 1s || N/A || 100% || 1.2 x 1e10/5s || ? || Bragg peak || ||
| 5 || f/7.1 || 1s || N/A || 100% || 1.2 x 1e10/5s || 80.5 || Bragg peak || No beam caught ||
||
||
||
||
|-
|-


| 6 || f/7.1 || 1s || N/A || 100% || 1.2 x 1e10/5s || ? || Bragg peak || ||
| 6 || f/7.1 || 1s || N/A || 100% || 1.2 x 1e10/5s || 80.5 || Bragg peak || No beam caught ||
||
||
||
||
|-
|-


| 7 || f/7.1 || 1s || N/A || 100% || 1.2 x 1e10/5s || ? || Bragg peak || ||
| 7 || f/7.1 || 1s || N/A || 100% || 1.2 x 1e10/5s || 80.5 || Bragg peak || Perfect for DSLR, saturating sensor ||
||
||
||
||
|-
|-


| 8 || f/5 || 1s || High || 50% || 0.6 x 1e10/5s || ? || Bragg peak ||  ||
| 8 || f/5 || 1s || High || 50% || 0.6 x 1e10/5s || 80.5 || Bragg peak || Good for sensor.77 mm of PMMA (11 sheets) in front of enclosure for calibration tests. ||
||
||
||
||
|-
|-


| 9 || f/5.6 || 1s || N/A || 50% || 0.6 x 1e10/5s || ? || Bragg peak ||  ||
| 9 || f/5.6 || 1s || N/A || 50% || 0.6 x 1e10/5s || 80.5 || Bragg peak ||  ||
||
||
||
||
Line 112: Line 112:
|-
|-


| 14 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 14 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 1 ||
||
||
||
||
|-
|-


| 15 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 15 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 2 ||
||
||
||
||
|-
|-


| 16 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 16 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 3 ||
||
||
||
||
|-
|-


| 17 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 17 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 4 ||
||
||
||
||
|-
|-


| 18 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 18 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 5 ||
||
||
||
||
|-
|-


| 19 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s ||  || Bragg peak || ||
| 19 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s ||  || Bragg peak || Spot 6 ||
||
||
||
||
|-
|-


| 20 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 20 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 7 ||
||
||
||
||
|-
|-


| 21 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 21 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 8 ||
||
||
||
||
|-
|-


| 22 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 22 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 9 ||
||
||
||
||
|-
|-


| 23 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 23 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 10 ||
||
||
||
||
|-
|-


| 24 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 24 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 11 ||
||
||
||
||
|-
|-


| 25 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 25 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 12 ||
||
||
||
||
|-
|-


| 26 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 26 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 13. Calibration spot for middle of scintillator that we align the beam to. ||
||
||
||
||
|-
|-


| 27 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 27 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 14 ||
||
||
||
||
|-
|-


| 28 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 28 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 15 ||
||
||
||
||
|-
|-


| 29 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 29 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 15. Repeat with beam with more intensity. ||
||
||
||
||
|-
|-


| 30 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Bragg peak || ||
| 30 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Bragg peak || PMMA removed! ||
||
||
||
||
Line 272: Line 272:
|-
|-


| 46 || colspan="2" | N/A || High || 50% || 0.6 x 1e10/5s || 78.7 || Energy Scan || ||
| 46 || colspan="2" | N/A || High || 50% || 0.6 x 1e10/5s || 78.7 || Energy Scan || Sensor saturation! Reduce intensity. ||
||
||
||
||
Line 357: Line 357:
|-
|-


| 63 || f/3.2 || 1s || Low || 10% || 1.2 x 1e9/5s || 85.6 || Intensity Scan || ||
| 63 || f/3.2 || 1s || Low || 10% || 1.2 x 1e9/5s || 85.6 || Intensity Scan || Sensor saturation! ||
||
||
||
||
|-
|-


| 64 || f/3.2 || 5s || High || 2.22% || 2.66 x 1e8/5s || 85.6 || Intensity Scan || ||
| 64 || f/3.2 || 5s || High || 2.22% || 2.66 x 1e8/5s || 85.6 || Intensity Scan || Factor 4.5 reduction ||
||
||
||
||
Line 372: Line 372:
|-
|-


| 66 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 66 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 1. 77 mm of PMMA (11 sheets) in front of enclosure for calibration tests. Acquire 2nd spill of every spot. ||
||
||
||
||
|-
|-


| 67 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan ||  ||
| 67 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 2. 5 - 6 spills per spot. Column K shows the number of the spill recorded, starting at 0 with an empty spill, 1 - 1st spill of 1st spot, 2 - 2ns spill of 1st spot etc. So 7 is 2nd spill of 2nd spot.
  ||
||
||
||
||
|-
|-


| 68 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 68 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 3 ||
||
||
||
||
|-
|-


| 69 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 69 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 4 ||
||
||
||
||
|-
|-


| 70 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 70 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 5 ||
||
||
||
||
|-
|-


| 71 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 71 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 6 ||
||
||
||
||
|-
|-


| 72 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 72 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 7 ||
||
||
||
||
|-
|-


| 73 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 73 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 8 ||
||
||
||
||
|-
|-


| 74 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 74 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 9 ||
||
||
||
||
|-
|-


| 75 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 75 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 10 ||
||
||
||
||
|-
|-


| 76 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 76 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 11 ||
||
||
||
||
|-
|-


| 77 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 77 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 12 ||
||
||
||
||
|-
|-


| 78 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 78 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 13 ||
||
||
||
||
|-
|-


| 79 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 79 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 14 ||
||
||
||
||
|-
|-


| 80 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || ||
| 80 || f/5.6 || 1s || High || 50% || 0.6 x 1e10/5s || 252.7 || Spot Scan || Spot 15 ||
||
||
||
||
|-
|-


| 81 || f/3.2 || 1s || High || 50% || 0.6 x 1e10/5s || 99.6 || Radiation damage assessment ||  ||
| 81 || f/3.2 || 1s || High || 50% || 0.6 x 1e10/5s || 99.6 || Radiation damage assessment || 56 mm of PMMA (8 sheets) in front of enclosure. In stack: 3 mm scintillator sheet and 2.6 mm sheet (from front of stack).
  ||
||
||
||
||
Line 457: Line 459:
|-
|-


| 83 || f/3.2 || 1s || High || 50% || 0.6 x 1e10/5s || 99.6 || Radiation damage assessment ||  ||
| 83 || f/3.2 || 1s || High || 50% || 0.6 x 1e10/5s || 99.6 || Radiation damage assessment ||  Note: no beam seen at 86.4 MeV ||
||
||
||
||

Revision as of 23:22, 23 March 2018

Beam tests of the UCL plastic scintillator range calorimeter.

Night of Thursday 22th to Friday 23rd of March 2018

Pre-data Taking Information and Notes

  • Data was acquired with the DSLR camera and the ISDI CMOS sensor.
  • The average background value per pixel for the sensor is ~ 2500. The sensor pixel dimensions are 1030 x 1536. The sensor frame rate was always set to 25 frames per second.
  • NO optical grease between scintillator and sensor.
  • Scintillator stack arrangement: 15 x 2 mm, 15 x 2.6 mm and 1 x 3 mm sheets used: In ascending order from the back of the scintillator to the front/beam. Except for sheet number 21 (2mm) which is placed at the end (away end from the beam) of the stack because of a crack in the sheet.
  • Folder for DSLR images: digiCamControl/20180322_medaustron
  • Note: When powering on PC, sensor has to be connected, but unpowered!
  • 16383 = saturation pixel number (16 bit x 1 tap)
  • Sensor images are always saved as a sequence of 21 Tiff images
  • Folder for sensor images: XCAP/data/20180322_medaustron


Data Taking

DSLR Sensor Beam Info Thumbnails
Run number Aperture Exposure (s) Full Well Mode Intensity (%) Number of particles/s Beam Energy (MeV) Test type Comments DSLR Sensor
1 f/5 1s Low N/A Background Background run with ND filter

DSC_0001.jpg

N/A
2 f/5 1s Low 100% 1.2 x 1e10/5s 80.5 Bragg peak ND filter in - not enough light passes through filter. Remove! DSLR slightly saturated. 10 spills used for this run.
3 f/8 1s High N/A Background No ND filter!
4 f/8 1s High 100% 1.2 x 1e10/5s 80.5 Bragg peak
5 f/7.1 1s N/A 100% 1.2 x 1e10/5s 80.5 Bragg peak No beam caught
6 f/7.1 1s N/A 100% 1.2 x 1e10/5s 80.5 Bragg peak No beam caught
7 f/7.1 1s N/A 100% 1.2 x 1e10/5s 80.5 Bragg peak Perfect for DSLR, saturating sensor
8 f/5 1s High 50% 0.6 x 1e10/5s 80.5 Bragg peak Good for sensor.77 mm of PMMA (11 sheets) in front of enclosure for calibration tests.
9 f/5.6 1s N/A 50% 0.6 x 1e10/5s 80.5 Bragg peak
10 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6 Bragg peak
11 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6 Bragg peak
12 f/5.6 1s High 50% 0.6 x 1e10/5s 62.4 Bragg peak
13 f/5.6 1s High 50% 0.6 x 1e10/5s 62.4 Bragg peak
14 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 1
15 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 2
16 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 3
17 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 4
18 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 5
19 f/5.6 1s High 50% 0.6 x 1e10/5s Bragg peak Spot 6
20 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 7
21 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 8
22 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 9
23 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 10
24 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 11
25 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 12
26 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 13. Calibration spot for middle of scintillator that we align the beam to.
27 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 14
28 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 15
29 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 15. Repeat with beam with more intensity.
30 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Bragg peak PMMA removed!
31 f/5.6 1s High 50% 0.6 x 1e10/5s 101.8 Energy Scan
32 f/5.6 1s High 50% 0.6 x 1e10/5s 99.6 Energy Scan
33 f/5.6 1s High 50% 0.6 x 1e10/5s 97.4 Energy Scan
34 f/5.6 1s High 50% 0.6 x 1e10/5s 95.1 Energy Scan
35 f/5.6 1s High 50% 0.6 x 1e10/5s 92.8 Energy Scan
36 f/5.6 1s High 50% 0.6 x 1e10/5s 90.4 Energy Scan
37 f/5.6 1s High 50% 0.6 x 1e10/5s 86.4 Energy Scan
38 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6 Energy Scan
39 f/5.6 1s High 50% 0.6 x 1e10/5s 84.7 Energy Scan
40 f/5.6 1s High 50% 0.6 x 1e10/5s 83.9 Energy Scan
41 f/5.6 1s High 50% 0.6 x 1e10/5s 83.0 Energy Scan
42 f/5.6 1s High 50% 0.6 x 1e10/5s 82.3 Energy Scan
43 f/5.6 1s High 50% 0.6 x 1e10/5s 81.3 Energy Scan
44 f/5.6 1s High 50% 0.6 x 1e10/5s 80.5 Energy Scan
45 f/5.6 1s High 50% 0.6 x 1e10/5s 79.6 Energy Scan
46 N/A High 50% 0.6 x 1e10/5s 78.7 Energy Scan Sensor saturation! Reduce intensity.
47 f/5.6 1s High 20% 2.4 x 1e9/5s 78.7 Energy Scan
48 f/5.6 1s High 20% 2.4 x 1e9/5s 77.8 Energy Scan
49 f/5 1s High 20% 2.4 x 1e9/5s 76.9 Energy Scan
50 f/5 1s High 20% 2.4 x 1e9/5s 76.0 Energy Scan
51 f/5 1s High 20% 2.4 x 1e9/5s 75.1 Energy Scan
52 f/5 1s High 20% 2.4 x 1e9/5s 74.2 Energy Scan
53 f/5 1s High 20% 2.4 x 1e9/5s 72.4 Energy Scan
54 f/5 1s High 20% 2.4 x 1e9/5s 70.5 Energy Scan
55 f/5 1s High 20% 2.4 x 1e9/5s 68.5 Energy Scan
56 f/5 1s High 20% 2.4 x 1e9/5s 66.5 Energy Scan
57 f/5 1s High 20% 2.4 x 1e9/5s 64.5 Energy Scan
58 f/5 1s High 20% 2.4 x 1e9/5s 62.4 Energy Scan
59 f/7.1 1s High 100% 1.2 x 1e10/5s 85.6 Intensity Scan
60 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6 Intensity Scank
61 f/4.5 1s High 20% 2.4 x 1e9/5s 85.6 Intensity Scan
62 f/3.2 1s High 10% 1.2 x 1e9/5s 85.6 Intensity Scan
63 f/3.2 1s Low 10% 1.2 x 1e9/5s 85.6 Intensity Scan Sensor saturation!
64 f/3.2 5s High 2.22% 2.66 x 1e8/5s 85.6 Intensity Scan Factor 4.5 reduction
65 f/3.2 5s Low 2.22% 2.66 x 1e8/5s 85.6 Intensity Scan
66 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 1. 77 mm of PMMA (11 sheets) in front of enclosure for calibration tests. Acquire 2nd spill of every spot.
67 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 2. 5 - 6 spills per spot. Column K shows the number of the spill recorded, starting at 0 with an empty spill, 1 - 1st spill of 1st spot, 2 - 2ns spill of 1st spot etc. So 7 is 2nd spill of 2nd spot.
68 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 3
69 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 4
70 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 5
71 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 6
72 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 7
73 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 8
74 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 9
75 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 10
76 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 11
77 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 12
78 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 13
79 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 14
80 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 15
81 f/3.2 1s High 50% 0.6 x 1e10/5s 99.6 Radiation damage assessment 56 mm of PMMA (8 sheets) in front of enclosure. In stack: 3 mm scintillator sheet and 2.6 mm sheet (from front of stack).
82 f/3.2 1s High 50% 0.6 x 1e10/5s 99.6 Radiation damage assessment
83 f/3.2 1s High 50% 0.6 x 1e10/5s 99.6 Radiation damage assessment Note: no beam seen at 86.4 MeV



End of range telescope tests.