Proton Calorimetry/Experimental Runs/2018/Mar22: Difference between revisions

From PBTWiki
Jump to navigation Jump to search
No edit summary
Line 21: Line 21:
[http://www.hep.ucl.ac.uk/pbt/wikiData/photos/MedAustron_22-03-18/IMG_3067.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/photos/MedAustron_22-03-18/IMG_3067.jpg]
[http://www.hep.ucl.ac.uk/pbt/wikiData/photos/MedAustron_22-03-18/IMG_3067.jpg http://www.hep.ucl.ac.uk/pbt/wikiData/photos/MedAustron_22-03-18/IMG_3067.jpg]
<br />
<br />
</div>  
</div>
 


=== Data Taking ===
=== Data Taking ===

Revision as of 00:02, 24 March 2018

Beam tests of the UCL plastic scintillator range calorimeter.

Night of Thursday 22th to Friday 23rd of March 2018

Pre-data Taking Information and Notes

  • Data was acquired with the DSLR camera and the ISDI CMOS sensor.
  • The average background value per pixel for the sensor is ~ 2500. The sensor pixel dimensions are 1030 x 1536. The sensor frame rate was always set to 25 frames per second.
  • NO optical grease between scintillator and sensor.
  • Beam is extracted over 5 seconds (one spill). The full capacity (100%) of the synchrotron ring is 1.2x1e10 protons.
  • Scintillator stack arrangement: 15 x 2 mm, 15 x 2.6 mm and 1 x 3 mm sheets used: In ascending order from the back of the scintillator to the front/beam. Except for sheet number 21 (2mm) which is placed at the end (away end from the beam) of the stack because of a crack in the sheet.
  • Folder for DSLR images: digiCamControl/20180322_medaustron
  • Note: When powering on PC, sensor has to be connected, but unpowered!
  • 16383 = saturation pixel number (16 bit x 1 tap)
  • Sensor images are always saved as a sequence of 21 Tiff images
  • Folder for sensor images: XCAP/data/20180322_medaustron

IMG_3067.jpg

Data Taking

DSLR Sensor Beam Info Thumbnails
Run number Aperture Exposure (s) Full Well Mode Intensity (%) Number of particles/s Beam Energy (MeV) Test type Comments DSLR Sensor
1 f/5 1s Low N/A Background Background run with ND filter

DSC_0001.jpg

N/A
2 f/5 1s Low 100% 1.2 x 1e10/5s 80.5 Bragg peak ND filter in - not enough light passes through filter. Remove! DSLR slightly saturated. 10 spills used for this run.
3 f/8 1s High N/A Background No ND filter!
4 f/8 1s High 100% 1.2 x 1e10/5s 80.5 Bragg peak
5 f/7.1 1s N/A 100% 1.2 x 1e10/5s 80.5 No beam caught
6 f/7.1 1s N/A 100% 1.2 x 1e10/5s 80.5 No beam caught
7 f/7.1 1s N/A 100% 1.2 x 1e10/5s 80.5 Perfect for DSLR, saturating sensor
8 f/5 1s High 50% 0.6 x 1e10/5s 80.5 Good for sensor.77 mm of PMMA (11 sheets) in front of enclosure for calibration tests.
9 f/5.6 1s N/A 50% 0.6 x 1e10/5s 80.5
10 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6
11 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6
12 f/5.6 1s High 50% 0.6 x 1e10/5s 62.4
13 f/5.6 1s High 50% 0.6 x 1e10/5s 62.4
14 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 1
15 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 2
16 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 3
17 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 4
18 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 5
19 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 6
20 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 7
21 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 8
22 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 9
23 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 10
24 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 11
25 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 12
26 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 13. Calibration spot for middle of scintillator that we align the beam to.
27 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 14
28 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 15
29 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 15. Repeat with beam with more intensity.
30 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Energy Scan PMMA removed!
31 f/5.6 1s High 50% 0.6 x 1e10/5s 101.8
32 f/5.6 1s High 50% 0.6 x 1e10/5s 99.6
33 f/5.6 1s High 50% 0.6 x 1e10/5s 97.4
34 f/5.6 1s High 50% 0.6 x 1e10/5s 95.1
35 f/5.6 1s High 50% 0.6 x 1e10/5s 92.8
36 f/5.6 1s High 50% 0.6 x 1e10/5s 90.4
37 f/5.6 1s High 50% 0.6 x 1e10/5s 86.4
38 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6
39 f/5.6 1s High 50% 0.6 x 1e10/5s 84.7
40 f/5.6 1s High 50% 0.6 x 1e10/5s 83.9
41 f/5.6 1s High 50% 0.6 x 1e10/5s 83.0
42 f/5.6 1s High 50% 0.6 x 1e10/5s 82.3
43 f/5.6 1s High 50% 0.6 x 1e10/5s 81.3
44 f/5.6 1s High 50% 0.6 x 1e10/5s 80.5
45 f/5.6 1s High 50% 0.6 x 1e10/5s 79.6
46 N/A High 50% 0.6 x 1e10/5s 78.7 Sensor saturation! Reduce intensity.
47 f/5.6 1s High 20% 2.4 x 1e9/5s 78.7
48 f/5.6 1s High 20% 2.4 x 1e9/5s 77.8
49 f/5 1s High 20% 2.4 x 1e9/5s 76.9
50 f/5 1s High 20% 2.4 x 1e9/5s 76.0
51 f/5 1s High 20% 2.4 x 1e9/5s 75.1
52 f/5 1s High 20% 2.4 x 1e9/5s 74.2
53 f/5 1s High 20% 2.4 x 1e9/5s 72.4
54 f/5 1s High 20% 2.4 x 1e9/5s 70.5
55 f/5 1s High 20% 2.4 x 1e9/5s 68.5
56 f/5 1s High 20% 2.4 x 1e9/5s 66.5
57 f/5 1s High 20% 2.4 x 1e9/5s 64.5
58 f/5 1s High 20% 2.4 x 1e9/5s 62.4
59 f/7.1 1s High 100% 1.2 x 1e10/5s 85.6 Intensity Scan
60 f/5.6 1s High 50% 0.6 x 1e10/5s 85.6
61 f/4.5 1s High 20% 2.4 x 1e9/5s 85.6
62 f/3.2 1s High 10% 1.2 x 1e9/5s 85.6
63 f/3.2 1s Low 10% 1.2 x 1e9/5s 85.6 Sensor saturation!
64 f/3.2 5s High 2.22% 2.66 x 1e8/5s 85.6 Factor 4.5 reduction
65 f/3.2 5s Low 2.22% 2.66 x 1e8/5s 85.6
66 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot Scan Spot 1. 77 mm of PMMA (11 sheets) in front of enclosure for calibration tests. Acquire 2nd spill of every spot.
67 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 2. 5 - 6 spills per spot. Column K shows the number of the spill recorded, starting at 0 with an empty spill, 1 - 1st spill of 1st spot, 2 - 2ns spill of 1st spot etc. So 7 is 2nd spill of 2nd spot.
68 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 3
69 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 4
70 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 5
71 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 6
72 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 7
73 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 8
74 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 9
75 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 10
76 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 11
77 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 12
78 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 13
79 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 14
80 f/5.6 1s High 50% 0.6 x 1e10/5s 252.7 Spot 15
81 f/3.2 1s High 50% 0.6 x 1e10/5s 99.6 Radiation damage assessment 56 mm of PMMA (8 sheets) in front of enclosure. In stack: 3 mm scintillator sheet and 2.6 mm sheet (from front of stack).
82 f/3.2 1s High 50% 0.6 x 1e10/5s 99.6
83 f/3.2 1s High 50% 0.6 x 1e10/5s 99.6 Note: no beam seen at 86.4 MeV



End of range telescope tests.