Radiotherapy/Experimental Runs/2020-02-04

From PBTWiki
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

Second experimental measurements with Cyberknife at Birmingham. Green scintillator thin sheet measurements only.

Setup

Schematic

setup.png

Pre-data Taking Information and Notes

  • Phantom
    • 200mmx200mmx2mm blue scintillator sheet polystyrene base
    • sandwiched by two 200mmx200mmx50mm polystyrene clear plastic blocks
  • 565mm from lens to sheet
  • Check camera in focus with ruler on the sheet
  • Ruler in focus at sheet
  • Scintillator blocks aligned to laser cross hairs
  • 785 mm source to surface distance suitable for output factor/PDD/profile measurements.
  • Wooden box is covered with black cloth
  • Camera height adjusted with wooden plates
  • Top surfaces of plastic blocks and sheet aligned with paper underneath

Data Taking

  • All images taken with
    • Canon EOS 1100D
    • 18-55mm lens
    • f/5
    • 200ISO
    • 30s exposure
  • Beam turned on for a given amount of time within the 30s exposure
  • This is noted in monitor units, with the beam having constant 1000MU/minute
  • Used EOS utility app for controlling camera
  • Image of ruler taken at sheet for pixels to mm conversion
  • 3 images taken at each collimator size with 1 background image each.


Collimator Size (mm) Monitor Units With sheet No sheet
5 400

calibration_098.JPG

calibration_166.JPG

7.5 400

calibration_101.JPG

calibration_163.JPG

10 400

calibration_104.JPG

calibration_160.JPG

12.5 400

calibration_107.JPG

calibration_157.JPG

15 400

calibration_110.JPG

calibration_153.JPG

20 400

calibration_113.JPG

calibration_150.JPG

25 300

calibration_116.JPG

calibration_147.JPG

30 300

calibration_119.JPG

calibration_144.JPG

35 250

calibration_122.JPG

calibration_141.JPG

40 250

calibration_125.JPG

calibration_138.JPG

50 150

calibration_128.JPG

calibration_135.JPG

60 150

calibration_095.JPG

calibration_132.JPG

Before and after irradiation

5mm_010.JPG calibration_092.JPG

Green discolouration observed after irradiation. Only present when the lights are on